If the transmitter or remote indication has not yet been correctly set and adjusted, the
remote display dashes.
- - - - -
5.2.2
Test menu
Press keys
and
to change to the test menu:
TEST
Press
to control all display elements.
Press key
for additional instrument tests.
Press key
several times to quit the test menu.
5.3 Device coniguration
5.3.1
Parameter selection
To set the device, push the LOCK switch into the open position
, see also
Chapter
Otherwise no coniguration is possible, but it is only possible to see the set parameters!
-
When pressing key
, the program changes to the OP mode and displays the irst
parameter oP 010; see also Chapter
.
-
Press key
to display the parameter name.
-
If key
is pressed again, the irst selectable option of this parameter is displayed.
-
Press keys
or
to select the parameter.
-
Press key
to go to the next parameter.
-
If a parameter was changed, "save" is displayed in the last parameter after pressing
key
.
-
After pressing key
, "yes" appears in the display.
-
You can switch between "yes" and "no" by pressing keys
or
.
-
By selecting "yes" and pressing key
, the changes are saved in EAROM.
-
By selecting "no" and pressing key
, the changes are canceled.
-
Press key
twice to quit the set-up menu.
Selecting a speciic parameter is also possible directly, i.e. independent of the predeined
order.
For this, the displayed parameter number must be changed and then the
key must
be pressed, see Chapters
and
.
5 Getting started
Remote display PR 5110/60, ../70
EN-23
Minebea Intec
Summary of Contents for PR 5110
Page 35: ......