NOTICE
Defective cable junction box due to penetration of water.
Special valve must not be removed.
6.4 Power supply
The power supply unit PR 6024/62S (see Chapter
supply of the load cells.
6.5 Data and supply cables
The PR 6124/xxD data cable must be used to connect the PR 6024/68S cable junction box
to the device, max. length: 300 m.
The PR 6124/xxP power cable must be used to connect the PR 6024/62S power supply to
the PR 6024/68S cable junction box, max. length: 200 m
Note:
For order numbers, see Chapter
6.6 Cable connections
Note:
All components are only shown schematically.
Color Code
rd
=
red
gn
=
green
bu
=
blue
gy
=
gray
NOTICE
The functionality of the load cells is also threatened by lightning strikes near the
scale.
Make sure to use cable junction box PR 6024/68S (see Chapter
For more information, see the cable junction box installation manual.
NOTICE
Possible destruction of the load cells
The digital load cell will be destroyed if connected incorrectly.
Make the connection as shown in the illustrations.
6 Connection
Pendeo® Process-Digital Precision Compression Load Cell PR 6204
Minebea Intec
EN-19