SP-L error
[22] (SP-L) is written in input register address 52 when the input load cell output value is
0 mV/V or less, which is less than the range of zero calibration on the negative side.
Condition
Error code
Contents
SP-L
0 x 10
3
SPAN point – Zero point ≤ 0.0 mV/V
SP-H error
[23] (SP-H) is written in input register address 52 when the read load cell output value is
more than 3.1 mV/V, which exceeds the range of zero calibration on the positive side.
Condition
Error code
Contents
SP-H
0 x 10
3
3.1 mV/V < Span point – Zero point
3.1 mV/V < SPAN point
Error outside the setting range
When data outside the setting range is entered, the following codes are set.
Condition
Error code
Contents
Illegal data
0 x 10
3
When data outside [1], [10], [–1] and[–10] is written.
17.1.2.6
Finish Zero point and Span point ine calibration
Updated data for Zero point and the Span point is registered temporarily is saved in
holding register address 56848. The device returns to weighing mode.
[1] is written in holding register address 56848.
[1] is automatically rewritten as [0] after writing is complete.
Error outside the setting range
When data outside the setting range is entered, the following codes are set.
Condition
Error code
Contents
Illegal data
0 x 10
3
When data outside [1] is written
17.1.2.7
Calibration force inish
When [0] is written in holding register address 56833 during calibration, calibration is
discontinued without saving the calibration data that is registered temporarily, and the
device returns to weighing mode.
17.1.3
Calibration mode for digital linearization
Each item of the low chart corresponds to each procedure described in Chapter
17 ModBus communication
Weight indicator CSD-903
Minebea Intec
EN-159