![Minebea Intec Compact Scale Puro Advanced EF-P1 Operating Instructions Manual Download Page 33](http://html1.mh-extra.com/html/minebea-intec/compact-scale-puro-advanced-ef-p1/compact-scale-puro-advanced-ef-p1_operating-instructions-manual_1801200033.webp)
The display shows the 1
st
linearization weight value (50% of scale capacity)with all
digit blinking [00075.0] (in this example, 75 kg). This value cannot be modified.
[--C--] will be displayed while zero reading is stored.
9. Put the 1
st
linearization weight on the pan.
10. Press
Off
(Yes) button to start linearity adjustment at 50% of scale capacity
The display shows the the 2
nd
linearization weight value at 100% of scale capacity
with all digit blinking [00150.0] (in this example, 150 kg).
[--C--] will be displayed while reading is stored.
11. Put 2
nd
linearization weight on the pan.
12. Press
Off
(Yes) button to start linearity adjustment at 100% of scale capacity.
[--C--] is displayed while reading is stored.
7 Calibration and Adjustment
Compact Scale Puro® - Basic
Minebea Intec
EN-31