Minebea Intec Combics CAPXS U Series Operating Instructions Manual Download Page 13

12

 

Combics CAPXS.U

 

Combics CAPXS.U

   

13

 

 
 
 
 
 
 
 
 
 
 
 
 
 

To verify the availability of the Approved product, please refer to

 

www.approvalguide.com

 

 

THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE 

 

FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA 
T: +1 (1) 781 762 4300   F: +1 (1) 781 762 9375  E-mail: 

[email protected]

  

www.fmapprovals.com

 

 

F 347 (Mar 16) 

Page 1 of 3 

   

    

  

CERTIFICATE OF CONFORMITY

  

 

 

1. 

HAZARDOUS (CLASSIFIED) LOCATION ELECTRICAL EQUIPMENT PER US REQUIREMENTS  

2. 

Certificate No: 

FM18US0227X 

3. 

Equipment: 
(Type Reference and Name) 
 

Model CAPX..U-……-……, IUX.4-……-…… and IFX.4-……-
….. Weighing Platforms and Model YDI05-Z Interface 
Converter 
 

4. 

Name of Listing Company: 
 

Minebea Intec Bouvenden GmbH & Co. KG 

5. 

Address of Listing Company: 

Leinetal 2 
37120, Bovenden 
Germany 
 

6. 

The examination and test results are recorded in confidential report number: 
 

3023378 dated 3

rd

 March 2006 

7. 

FM Approvals LLC, certifies that the equipment described has been found to comply with the following Approval 
standards and other documents: 

 

FM Class 3600:2018, FM Class 3610:2010, FM Class 3611:2018, FM Class 3810:2018, 

ANSI/ISA 60079-0:2009, ANSI/UL 60079-11:2009 

 

8. 

If the sign ‘X’ is placed after the certificate number, it indicates that the equipment is subject to specific conditions 
of use specified in the schedule to this certificate. 
 

9. 

This certificate relates to the design, examination and testing of the products specified herein. The FM Approvals 
surveillance  audit  program  has  further  determined  that  the  manufacturing  processes  and  quality  control 
procedures in place are satisfactory to manufacture the product as examined, tested and Approved. 
 

10.  Equipment Ratings: 

 

For CAPX..U-...-..., IUX.4-…-..., IFX.4-...-... Weighing Platforms

.

 

 
Intrinsically Safe circuits or connections for Class I, II, III,  Division 1, Groups A, B, C, D, E, F and G T4 at Ta* 
when connected per 35739-003-07-A4; Class I Zone 1, AEx ia IIC T4 at  Ta*  when connected per 35739-003-07-

Certificate issued by: 

 

 

 

 

 
 
11 November 2018 

J. E. Marquedant 
VP, Manager, Electrical Systems 

Date 

 

 

SCHEDULE

 

 

US Certificate Of Conformity No: FM18US0227X  

 

THIS CERTIFICATE MAY ONLY BE REPRODUCED IN ITS ENTIRETY AND WITHOUT CHANGE 

 

FM Approvals LLC. 1151 Boston-Providence Turnpike, Norwood, MA 02062 USA 
T: +1 (1) 781 762 4300   F: +1 (1) 781 762 9375  E-mail: 

[email protected]

  

www.fmapprovals.com

 

 
F 347 (Mar 16) 

Page 2 of 3 

A4, Class I, Division 2, Groups A, B, C and D T4 at Max. Ta=50°C;  Class II, III, Division 2, Groups E, F and G T4 
at  Ta=50°C and Class II, Zone 20, IIIC T4 at Ta=* when connected as per control drawings 35739-003-07-A4. 
 

*see table below 

V

max

 

I

max 

P

C

i

 (cable)  L

(cable)

 

T

17Vdc  410mA 

1.25W  162pF/m 

0,6µH/m 

60°C 

13Vdc  410mA 

2.00W  162pF/m 

0,6µH/m 

40°C 

 

For YDI05-Z.. Interface Converter. 

 

Associated Intrinsically Safe for Class I, II, III, Divisions 1 and 2, Groups A, B, C, D, E, F  and G as per control 
drawing 65710-800-07-A4; Entity;  Class II, [Zone 20], [Ex ia] IIIC as per control drawing 65710-800-07-A4; Entity 
Class I, [Zone 0], [Ex ia] IIC as per control drawing 65710-800-07-A4; Entity 
 

11.  The marking of the equipment shall include: 

 

 

For Models CAPX, IUX.4 and IFX.4 

Class I, II, III, Division 1, Groups A, B, C, D, E, F and G; T4 Ta = *  – 35739-003-07-A4; Entity 

Class I, Division 2, Groups A, B, C, D; T4 Ta = +50°C 

Class II, III, Division 2, Groups E, F, G; T4 Ta = +50°C  

Class II, [Zone 20], [AEx ia] IIIC T4 Ta = * – 35739-003-07-A4;  

Class I, Zone 0, AEx ia IIC T4 Ta* – 35739-003-07-A4; Entity 

V

max

 

I

max 

P

C

i

 (cable)  L

(cable)

 

T

17Vdc  410mA 

1.25W  162pf/m 

0,6µH/m 

60°C 

13Vdc  410mA 

2.00W  162pf/m 

0,6µH/m 

40°C 

 

For Model YDI05-Z 

Associated Intrinsically Safe Equipment for: 

Class I, II, III, Division 1,  Groups A, B, C, D, E, F and G –65710-800-07-A4;  Entity 

Zone [20], AEx [ib] IIIC – 65710-800-07-A4;  Entity 

Class [I], Zone [1],AEx [ia] IIC–65710-800-07-A4; Entity

 

 

 

 

12. 

Description of Equipment:

 

 

The Model CAPX..U-……-……, IUX.4-……-…… and IFX.4-……-….. are stainless steel bench and floor Weighing 
Platforms based on strain-gauge load cells. They can be connected to any indicator that fulfils the requirements of 
the desired type of protection. The load cells within the weighing platform are powered by the indicator which also 
measures the change in resistance within the load cell corresponding to the load that is applied onto the weighing 
platform. 
 
The Model YDI05-Z. is an intrinsically safe barrier and interface converter. It is used to connect devices located in 

Summary of Contents for Combics CAPXS U Series

Page 1: ...Operating Instructions Combics Minebea Intec Models CAPXS U Stainless Steel Weighing Platforms for Use in Hazardous Locations 98648 014 07 98648 014 07...

Page 2: ...ety information in its entirety during installation and operation as well as while performing maintenance and repair work on the equipment It is important that all personnel using the Combics equipmen...

Page 3: ...ebea Intec dealer or the Minebea Intec Service Center Any installation work that does not conform to the instructions in this man ual will result in forfeiture of all claims under the manufacturer s w...

Page 4: ...ence when the equip ment is operated in a commercial envi ronment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction...

Page 5: ...e leveling feet turn clockwise To lower the weighing platform retract the leveling feet turn counterclockwise After leveling the weighing platform retighten the locknuts securely as described below Lo...

Page 6: ...ontrol Drawing These parameters Entity Concept must be taken into account when connecting an indicator to the platform The explosion protected weighing system must be installed in accordance with ackn...

Page 7: ...ontact with the clamps 2 Expose approximately 15 cm 6 inches of the wires 3 for installation Attach the cable to the weighing platform as follows Expose approximately 5 cm 2 inches of the wires for in...

Page 8: ...HING 3 OMBICS STAINLESS STEEL 0LATFORM IMENSIONS IN INCHES UMBER OF LOAD CELLS 3ELECTION OF THE WEIGHING CAPACITY IN LBS You should not place loads on the scale that exceed its maximum weighing capaci...

Page 9: ...e use of steel wool sponges is not permitted Solvents are permitted for use only on stainless steel parts 10 10 Cleaning Unplug the scale from the AC power before cleaning To clean the weighing platfo...

Page 10: ...4 30 24 YAR CWSS 30 30 30 YAR CWSS 30 36 36 YAR CWSS 36 48 48 YAR CWSS 48 60 48 YAR CWSS 48 60 48 YAR CWSS 60 60 60 YAR CWSS 60 11 Drive on ramp AISI 316 Ti stainless steel tread plate for platform si...

Page 11: ...essory no DD 12 12 height 24 18 Adjustable to a max height of 25 31 YWT01CWP EE 18 18 height 24 18 Adjustable to a max height of 25 31 YWT02CWP Bench stainless steel for size Size in inch Accessory no...

Page 12: ...he equipment described has been found to comply with the following Approval standards and other documents FM Class 3600 2018 FM Class 3610 2010 FM Class 3611 2018 FM Class 3810 2018 ANSI ISA 60079 0 2...

Page 13: ...For Models CAPX IUX 4 and IFX 4 Class I II III Division 1 Groups A B C D E F and G T4 Ta 35739 003 07 A4 Entity Class I Division 2 Groups A B C D T4 Ta 50 C Class II III Division 2 Groups E F G T4 Ta...

Page 14: ...IFX 4 Weighing Platforms Model YDI05 Z Interface Converter 13 Specific Conditions of Use 14 Test and Assessment Procedure and Conditions This Certificate has been issued in accordance with FM Approval...

Page 15: ...as been found to comply with the following Approval standards and other documents CAN CSA C22 2 No 157 92 2006 CAN CSA C22 2 No 213 208 CAN CSA C22 2 No 142 2004 CAN CSA C22 2 No 60079 0 2007 CAN CSA...

Page 16: ...II III Divisions 1 and 2 Groups A B C D E F and G as per control drawing 65710 800 07 A4 Entity Zone 20 Ex ia IIIC as per control drawing 65710 800 07 A4 Entity Zone 1 Ex ia IIC as per control drawin...

Page 17: ...ntrinsically safe barrier and interface converter It is used to connect devices located in the non hazardous area with devices located in hazardous locations The signals that are exchanged between non...

Page 18: ...rnpike Norwood MA 02062 USA T 1 1 781 762 4300 F 1 1 781 762 9375 E mail information fmapprovals com www fmapprovals com F 348 Mar 16 Page 4 of 4 10th March 2006 Original Issue 11th November 2018 Supp...

Page 19: ...Combics CAPXS U 19...

Page 20: ...s than 1 ohm In Canada The Apparatus must be connected to a suitable ground electrode per Canadian Electrical Code Part 1 The resistance of the ground pad must be less than 1 ohm 3 The Entity Concept...

Page 21: ...cal Code NFPA 70 Article 504 or 505 The resistance of the ground pad must be less than 1 ohm In Canada The Apparatus must be connected to a suitable ground electrode per Canadian Electrical Code Part...

Page 22: ...22 Combics CAPXS U...

Page 23: ...Combics CAPXS U 23...

Page 24: ...means without prior written permission from Minebea Intec All rights reserved The status of the information specifications and illustrations in this manual is indicated by the date given below Minebe...

Reviews: