background image

P/N: 0002-08-50038-00 Rev A

24

 

Quick Reference Guide

V

V
V

Adjust Arrhythmia Alarm Responses

To adjust a patient’s alarm responses on the V12 or V21 bedside monitor:
  a. Select 

Alarms

 tab from Navigation Area

  b. Select 

Arrhythmia

 to enter the 

Arrhythmia

 tab

  c. Select 

Responses

 tab

  d.  Select arrhythmia 
  e.   Select 

• 

Audio On

 – to enable the audio notification. 

    •  

Print on Alarm

 – to activate an automatic printing when the alarm 

occurs

    •  

Save to Event

 – to save a 20 second event in the Events tab  

(

History 

menu) 

(A   in the box indicates selected)

  f. Select 

Accept

 to save settings and return to the main screen

Panorama® Central Station interface:

 Adjusting arrhythmia alarms at the 

Panorama Central Station or WorkStation 

will not

 change arrhythmia alarms 

on the V12 or V21 bedside monitor. Adjusting arrhythmia alarms at the V12 or 
V21 bedside monitor 

will

 affect the arrhythmia alarm on the Panorama Central 

Station, ViewStation, and WorkStation.

Disable Non-Lethal Arrhythmias 

To disable all arrhythmia alarms 

except lethal

 arrhythmias on the V12 or V21 

bedside monitor:
  a.  Select ECG waveform to enter 

ECG Setup

 tab

  b. Select 

Arrhythmia

 tab

  c. Select 

Lethals Only On

  d. Select 

Accept

 to save settings and return to the main screen

Or
  a. Select 

Alarms

 tab from Navigation Area

  b. Select 

Arrhythmia

 to enter the 

Arrhythmia

 tab

  c. Select 

Lethals Only On

  d. Select 

Accept

 to save settings and return to the main screen

Summary of Contents for V Series

Page 1: ...Quick Reference Guide V V ...

Page 2: ...intended as a replacement to the Operating Instructions Prior to operating equipment the user must be familiar with the Operating Instructions Manual contents This document is a guideline only to be used as an aid to comprehensive Inservice Training ...

Page 3: ... Monitoring 10 Privacy Mode 11 Print 11 Real Time Waveforms 11 Adjust Print Location 12 Adjust Waveforms Printing 12 ECG 13 Adjust ECG Lead 13 Enable Pacemaker Detection 13 Adjust Arrhythmia and Beat Classification Leads 14 Adjust QRS Threshold 14 Initiate Relearn 15 Initiate QT Monitoring 15 Enable QTC Limit 15 Initiate ST Analysis 16 12 Lead ECG 16 Enable 12 Lead ECG 16 View 12 Lead ECG 16 Print...

Page 4: ...0 Adjust Temperature Units 21 Alarms 21 Adjust Alarm Limits 21 Adjust Alarm Responses 22 Adjust Arrhythmia Alarms 23 Adjust Arrhythmia Alarm Responses 24 Disable Non Lethal Arrhythmias 24 Reset Alarms Version 2 5 or greater 25 Pause All Alarms Version 2 5 or greater 25 Silence Alarms Version 2 4 or earlier 26 Silence All Alarms Version 2 4 or earlier 26 Transfer 26 Hemodynamic Monitoring 27 Initia...

Page 5: ...y Selection 29 Adjust Display Selection 29 History 30 List Trends 30 Alarm Events 31 Full Disclosure 31 Volumes 32 Beat Tone 32 Date and Time 33 Parameter Colors 33 Default Preset Settings 33 Lock Unlocking Touchscreen Display 34 Table of Contents cont d ...

Page 6: ...r Events ECG Menu Full Disclosure Graphic Trends Hemodynamics Lead Size List Trends NIBP Stat PA Insertion Parameters PAWP Print 12 Lead Privacy Mode Quick Trends Relearn Arrhythmia Relearn ST Remote View Standby Start NIBP ST Review Transfer Discharge View All ECG Zero All NIBP Quick Functions Keys Configurable Up to 6 keys may be used in Quick Functions tab ...

Page 7: ...nt request Capture Event Saves an event to the event database History Accesses the Trends Events and Full Disclosure dialogs Silence All Silences all current and future alarms for the configured time period Silence Silences the current alarms for the configured time period Normal Screen Closes all dialogs and returns to the Main Screen Quick Functions Tab Configurable Start Stop NIBP Starts Stops ...

Page 8: ...ab Calculators Accesses the Calculators dialog Alarm Watch Accesses the Alarm Watch dialog Remote View Accesses the Remote View dialog System Information Accesses the System Information dialog VAccess Accesses the VAccess dialog Procedures Tab Note These buttons are only available when using the Adult or Pediatric patient size C O Run Accesses the Cardiac Output Runs dialog PAWP Accesses the PAWP ...

Page 9: ...on automatically enters it on the V12 or V21 bedside monitor Entering demographic information on the V12 or V21 bedside monitor automatically enters it on the Panorama Central Station ViewStation and WorkStation Entering Demographics is not required for patient information to be displayed or stored Adjust Patient Size The Patient Info menu allows the clinician to adjust patient size To adjust pati...

Page 10: ...ient on the V12 or V21 bedside monitor automatically discharges the patient on the Panorama Central Station WorkStation and ViewStation Discharging the patient on the Panorama Central Station discharges the patient on the WorkStation ViewStation and V12 or V21 bedside monitor optional STANDBY Enter Standby The Standby function allows the clinician to suspend monitoring on a patient While in standb...

Page 11: ... turn off the display on the V12 or V21 bedside monitor when admitted to the Panorama Central Station a Select Privacy Mode Quick Function tab Or a Select Display Setup tab and select Privacy Mode To remove a patient from Privacy Mode and resume monitoring on the V12 or V21 bedside monitor touch anywhere on the screen Note If communication is interrupted between the V12 or V21 bedside monitor and ...

Page 12: ...er to print to a predetermined laser printer 3 Laser Thermal to simultaneously print to the modular recorder and a predetermined laser printer e Select Accept to save settings and return to the main screen Adjust Waveforms Printing To adjust printed waveforms on a patient on the V12 or V21 bedside monitor a Select Print Setup Print tab b Select Setup c Select Configuration Waveform 1 Select wavefo...

Page 13: ...dside monitor Panorama Central Station WorkStation and ViewStation are independent Enable Pacemaker Detection The V12 or V21 bedside monitor allows the clinician to enable pacemaker enhancement To enable the pacemaker enhancement on the V12 or V21 bedside monitor a Select ECG waveform to enter ECG Setup tab Or a Select Parameters Setup tab and select ECG to enter ECG Setup tab b Select Pacer tab c...

Page 14: ...condary Analysis Lead f Scroll and select lead from drop down menu g Select Beat Classification Lead h Scroll and select lead from drop down menu i Select Accept to save settings and select Done as needed to return to the main screen Adjust QRS Threshold Version 2 5 or greater The V12 or V21 bedside monitor allows the clinician to manually adjust the QRS threshold on detection to avoid noise spike...

Page 15: ... a Select ECG waveform to enter ECG Setup tab Or a Select Parameters Setup tab and select ECG to enter ECG Setup tab b Select Advanced Arrhy Setup tab c Select QT Monitoring and select On d Select Accept to save settings and select Done as needed to return to the main screen Enable QTC Limit Version 2 5 or greater a Select ECG waveform to enter ECG Setup tab Or a Select Parameters Setup tab and se...

Page 16: ...o enable 12 Lead ECG on V12 or V21 bedside monitor a Prep patient s skin in accordance with hospital policy b Attach electrode to lead wire before placing on patient c Place electrodes on patient in accordance with hospital policy d Insert 12 Lead ECG Module into V Patient Server VPS e Select ECG waveform to enter ECG Setup tab f Select ECG Cable Type g Scroll and select 12 Lead from drop down men...

Page 17: ... b Select Print 12 Lead to print to the configured location Or a Select Procedures tab in Navigation Area b Select 12 Lead c Select Print 12 Lead to print to the configured location d Select Accept to return to the main screen NIBP Initiate a Non Invasive Blood Pressure Measurement NIBP To initiate a NIBP measurement on a patient on the V12 or V21 bedside monitor a Select on the VPS module Or a Se...

Page 18: ...Scroll and select interval e Select Accept to save settings and to return to the main screen Note When a NIBP interval is first selected Start NIBP must be selected If the NIBP interval is adjusted Start NIBP is not required Start NIBP must be re selected if the NIBP interval is interrupted i e Standby NIBP Stop etc During NIBP cuff inflation to stop all NIBP measurements on a patient on the V12 o...

Page 19: ...to enter the Invasive Blood Pressure s Setup tab c Select Label d Select label e Select Accept to save settings and select Done as needed to return to the main screen Note If the label is assigned to another IBP port even if the IBP is not being displayed or utilized the label will not be included in the list selection Zero Invasive Blood Pressure To zero one or more invasive blood pressure transd...

Page 20: ... main screen b Select preset s name from Display Preset list c Select Temporary Edit d Select 1st invasive blood pressure to overlap e Select Overlap f Select 2nd invasive blood pressure to overlap from drop down list g Select Accept to save settings and return to the Display Setup tab h Select Accept to return to the main screen TEMPERATURE The V12 or V21 bedside monitors allow a clinician to mon...

Page 21: ...ain screen Or a Select Parameters Setup tab and select Temperature to enter the Temperatures Setup tab b Select Units c Select C or F d Select Accept to save settings and select Done as needed to return to the main screen ALARMS Adjust Alarm Limits The Alarms tab allows the clinician to configure alarm limit and alarm responses on the V12 or V21 bedside monitor To adjust a patient s alarm limit on...

Page 22: ...mits on the V12 or V21 bedside monitor Changing alarm limits at the V12 or V21 bedside monitor will automatically change alarm limits on the Panorama Central Station and WorkStation Adjust Alarm Responses To adjust a patient s alarm responses on the V12 or V21 bedside monitor a Select Alarms tab from Navigation Area b Select Responses c Select applicable alarm limit group s tab i e ECG Resp Oximet...

Page 23: ... to enter ECG Setup tab b Select Arrhythmia tab c Select individual arrhythmia d Select On Off to enable or disable arrhythmia alarm notification A in the box indicates selected High Red Medium Yellow or Low Blue to adjust the audio and visual alarm indicators e Select Accept to save settings and return to the main screen Or a Select Alarms tab from Navigation Area b Select Arrhythmia to enter the...

Page 24: ...rama Central Station interface Adjusting arrhythmia alarms at the Panorama Central Station or WorkStation will not change arrhythmia alarms on the V12 or V21 bedside monitor Adjusting arrhythmia alarms at the V12 or V21 bedside monitor will affect the arrhythmia alarm on the Panorama Central Station ViewStation and WorkStation Disable Non Lethal Arrhythmias To disable all arrhythmia alarms except ...

Page 25: ...or V21 bedside monitor to silence all current alarms for a preconfigured time interval or until the alarm condition is resolved Panorama Central Station interface Selecting Alarm Acknowledge on the V12 or V21 bedside monitor will silence an alarm on the Panorama Central Station WorkStation and ViewStation for a preconfigured amount of time Selecting Mute at the Panorama Central Station will not si...

Page 26: ...erval Panorama Central Station interface Selecting Silence All on the V12 or V21 bedside monitor will silence active alarms on the Panorama Central Station WorkStation and ViewStation for a preconfigured amount of time Selecting Mute at the Panorama Central Station will not silence alarms on the V12 or V21 bedside monitor TRANSFER The V12 or V21 bedside monitor has the ability to transfer patient ...

Page 27: ...entricle each minute expressed in liters per minute l min Cardiac Index CI is the Cardiac Output divided by the patient s body surface area To initiate a Cardiac Output measurement on the V12 or V21 bedside monitor a Insert C O Module b Connect the C O cable to the C O connector on the monitor interconnecting the C O Module PA Catheter and injectate c Select C O tile to enter C O Alarms tab and se...

Page 28: ... when the PA distal balloon is inflated and the catheter advances and occludes a distal pulmonary artery PAWP pressure is a reflection of the pressure in the left ventricle at end diastole To initiate a PAWP measurement on the V12 or V21 bedside monitor a Select PA waveform or numeric tile and select PAWP Or a Select Procedures tab from the Navigation Area and select PAWP b Select Setup to adjust ...

Page 29: ...ally enter indicates manual entry c Select Calculate to complete calculation d Select Print to print the currently displayed Hemodynamic Calculations e Select Done to return to the main screen SCREEN DISPLAY The V12 and V21 allow the clinician to adjust the main display based on patient acuity Display Selection To enter the display selection menu and adjust the main screen s layout on the V12 or V...

Page 30: ... Digital and waveform display selections on the Panorama Central Station WorkStation and ViewStation are independent of the V12 or V21 bedside monitor HISTORY List Trends The List tab stores 3000 trend sets on a patient To display and adjust List trends on a patient on the V12 or V21 bedside monitor a Select History in the Navigation Area b Select List Trends tab c Select Display Group to specify ...

Page 31: ...nd print an alarm event on a patient on the V12 or V21 bedside monitor a Select History in the Navigation Area b Select Events tab e Scroll and select event f Select View and Detail to view the 20 second event g Select Print to print to the configured location h Select Done to return to the main screen Note The waveforms printed are determined by Setup in the Event or Disclosure tab on the V12 or ...

Page 32: ...Setup tab from Navigation area b Select Audio Levels c Select volume ranging from X 10 X minimum 10 max using arrows on Alarm Volume d Select Accept to return to the main screen Note The minimum volume range is predetermined during installation Version 2 5 or greater BEAT TONE To adjust the beat tone QRS tone volume on the V12 or V21 bedside monitor a Select Setup tab from Navigation area b Select...

Page 33: ...V12 or V21 bedside monitor The date and time cannot be adjusted at the V12 or V21 bedside monitor when interfacing with the Panorama Central Station PARAMETER COLORS The V12 or V21 bedside monitor provides the ability to adjust the parameter s numeric and waveform if applicable color To adjust a parameter s color on the V12 or V21 bedside monitor a Select parameter s waveform or numeric tile b Sel...

Page 34: ...ck the touch screen display temporarily for cleaning purposes To temporarily lock the display on the V12 or V21 bedside monitor a Select Setup tab in Navigation area b Select Display to enter Display Options tab c Select Screen Lock d A 10 second screen lock is displayed allowing for cleaning of the screen e Select Accept to return to the main screen ...

Page 35: ...Mindray DS USA Inc 800 MacArthur Blvd Mahwah NJ 07430 Tel 800 288 2121 or 201 995 8000 www mindray com P N 0002 08 50038 00 Rev A ...

Reviews: