27 - 5
27.4.3
The Latching Tab
27.4.4
The Remote View Tab
Menu Item
Default Setting
Description
Lethal
Visible
Unselected
Selects alarm latching rules:
•
If
Visible
is selected, you can separately latch
visual alarm signal.
•
Latching audible alarm signal simultaneously
latches visual signal.
•
Selecting alarms of lower priority
simultaneously latches higher priority alarms.
Audible
High
Visible
Audible
Med Visible
Audible
Low
Visible
Audible
Menu Item
Default Setting
Description
Reset Remote Bed’s Alarms
Off
Selects whether you can reset alarms occurring
to the remote devices from your monitor.
On
: the
Alarm Reset
button appears on the
bottom left of the
Remote View
screen.
Alarm Reset By Other Bed
On
On
: alarms on your monitor can be reset by
remote devices.
Alarm Reminder
Audible
Selects what alarm indicators are necessary for
the remote devices.
•
V Audible
: the monitor provides
visual alarm indication, and continuous
audible alarm indication if the alarm persists
at the remote device.
•
V Single Tone
: the monitor provides
visual alarm indication, and a single tone
when the alarm occurs at the remote device.
•
Visible Only
: the monitor only provides visual
alarm indication.
Alarm Priority
All
Selects what priority of remote device alarms are
presented for audible notification
•
All
: the monitor sounds if an alarm occurs.
•
High & Med
: the monitor sounds if a high or
medium priority alarm occurs.
•
High Only
: the monitor sounds only if a high
priority alarm occurs.
Alarm Sound
ISO
Selects the alarm tone pattern for the remote
device alarms.
Remote Disconnected Alarm
On
Selects whether an alarm is issued if a remote
device is disconnected.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...