Duo™ Service Manual
0070-10-0604-01
4 - 7
Repair Information
Disassembly Instructions
4.3.7
Power Supply Removal
1.
Remove the battery cable from P1 on the power supply PCB.
2.
Remove four (4) 3 x 6 mm Phillips panhead machine screws from the corners of the
power supply PCB.
3.
Remove one (1) 3 x 6 mm Phillips panhead machine screw from the line power entry
connector.
4.
Slide the Power Supply pcb towards the bottom (open end) of the battery housing frame
until it can go no farther.
5.
Lift the power supply PCB from the battery housing frame. Use care to avoid breaking
the positioning tab from the line power entry connector.