NIBP Calibration
Calibration and Performance Verification
2 - 6
0070-10-0604-01
Duo™ Service Manual
2.6
NIBP Calibration
2.6.1
Test Equipment
• NIBP simulator
• NIBP test chamber/dummy cuff
• Manometer with bulb
2.6.2
Test Procedure
2.6.2.1
Transducer Accuracy
1.
Connect the 500 cc Test Chamber and calibrated manometer via a ‘‘T’’ fitting to the
NIBP fitting on the
Duo
monitor under test.
2.
Ensure the
Duo
is not turned on. Simultaneously, press and hold the
POWER
and
NIBP
START
/
STOP
buttons.
When the monitor enters the NIBP Calibration Mode, message code 525 will be
displayed in the Information Codes window. Release the buttons simultaneously.
3.
Momentarily press the
NIBP START/STOP
button to start the NIBP calibration. Vent the
Test Chamber and verify the
Duo
and the manometer read zero. Using the bulb,
pressurize the test chamber to 50 mmHg and verify the
Duo
reading agrees with the
man/- 3 mmHg. Using the bulb, increase the pressure to 200 mmHg and verify
the
Duo
reading agrees with the man/- 3 mmHg.
2.6.2.2
Pneumatic Leak Test
1.
Connect the 500 cc test chamber to the NIBP fitting on the
Duo
monitor under test.
2.
From the NIBP Calibration Mode (code 525) momentarily press the
CLEAR
button on
the
Duo
keypad. The
Duo
will then switch to the Pneumatic Test Mode and will display
message code 550 in the Information Codes window.
3.
Momentarily press the
NIBP START/STOP
button to start the leak test. The
Duo
under
test will automatically pressurize the test chamber to approximately 180 mmHg.
4.
After approximately 20 seconds, the
Duo
under test will vent the pressure in the test
chamber and display a message code E06 (Pass) or E07 (Fail) in the systolic window.
2.6.2.3
Dynamic Repeatability Test
1.
Restart unit and allow it to enter normal operating mode.
2.
Use polyurethane tubing to connect the
Duo
monitor to a calibrated NIBP simulator and
the 500 cc test chamber/dummy cuff via a ‘‘T’’ fitting.
3.
Select Adult patient size for both the
Duo
under test and the NIBP simulator.
4.
Select a target simulated blood pressure within the ‘‘normal’’ range on the simulator.
5.
Take 10 successive NIBP readings and compare the systolic, diastolic, mean and heart
rate readings for consistency. Readings should not deviate more than +/- 5 mmHg for
the NIBP readings and +/- 2 bpm or 2%, whichever is greater for heart rate.