P/N: H-046-007060-00 Rev 3.0
11
BeneVision TD60
Quick Reference Guide
RESP
(OPTIONAL)
Set NCM Alarm Delay
The TD60 allows configuring the desired time for
NCM Alarm Delay
. When
the patient has stopped breathing for longer than the set
NCM Alarm Delay
time, the TD60 will trigger the "No Chest Movement" alarm.
To set
NCM Alarm Delay
:
•
On the main screen, tap the Resp digital area or Resp waveform area to
enter the
Resp
menu.
OR
1. Press to enter the main menu.
2. Tap
Parameter Setup
.
3. Tap
Resp
.
4. Set
NCM Alarm Delay
.
Note:
The "No Chest Movement" alarm is displayed on the Central Monitoring
System only.
Note:
The Resp functionality for the TMS60 telemetry monitoring system is
available at the CMS whose version is 03.00 and above only.