These guidelines do 

NOT

 replace the instructions for use and all users should refer to the appropriate 

Operator’s Manual

 for detailed instructions. 

The Illustrations may not necessarily reflect the setup or data displayed on your device. 

 

CAS BeneVision N Series 22/19 & N1 User Guide_V1     Mindray UK © 2020. Errors and omissions excepted. Subject to change. 

Mindray® is a registered trademark of Shenzhen Mindray Bio-Medical Electronics Co. Ltd. 

ALARMS 

 

 

Setting alarm limits 

 

Select the Alarm quick key 

OR 

 

Select Main Menu quick key 

 Setup from the Alarm column 

OR 

 

Individual alarm settings can also be accessed from each parameter   menu 

 

Then select desired buttons to set alarm   properties 

 

 

 
 

Alarm Reset 

 

 

 

 

The alarm sound is silenced 

 

A √ appears before the alarm message, indicating that the alarm is acknowledged 

 

The colour of the parameter numeric background corresponds with the alarm priority, but the parameter 

numeric does not flash 

 

 
 
 

Alarm Pause / Audio Pause 

Depending on system setup 

 

 

Alarm Pause quick key will temporarily disable all alarm indicators for a preset length of  time 

 

 

 

When alarms are paused, the following rules occur: 

 

No physiological alarm will be presented 

 

For technical alarms, alarm sounds are paused, but alarm lamps and alarm messages remain 

presented 

 

The remaining alarm pause time is displayed in the physiological alarm information area 

 

The alarm pause symbol is displayed in the system information area 

 

 

 

Audio Pause quick key pauses the alarm tone only for a preset length of time 

 

 

 
 

Please Note: 

When the alarm pause/audio pause time expires, the alarm/audio paused status is automatically 

deactivated. You can also cancel the status by pressing the Alarm Pause/Audio Pause quick key 