These guidelines do
NOT
replace the instructions for use and all users should refer to the appropriate
Operator’s Manual
for detailed instructions.
The Illustrations may not necessarily reflect the setup or data displayed on your device.
CAS BeneVision N Series 22/19 & N1 User Guide_V1 Mindray UK © 2020. Errors and omissions excepted. Subject to change.
Mindray® is a registered trademark of Shenzhen Mindray Bio-Medical Electronics Co. Ltd.
3
NIBP
Starting & Stopping NIBP measurements
•
Start and stop NIBP measurement by selecting one of the options below:
IBP
Changing invasive blood pressure labels
•
Select the IBP numeric area or waveform area to enter the corresponding pressure menu
→
Setup
→
Set IBP1
or IBP2 Label
Please Note:
The pressure label is a unique identifier for each type of pressure, therefore, you should select a
proper pressure label for the source of the pressure you want to monitor
Zeroing pressures
Zero the transducer by one of the following methods:
•
Press the Zero hard key on the module
•
Select the numeric area (such as the Art numeric area), and then select Zero option
•
Select the IBP Zero quick key
Please Note:
After the zero calibration is completed, close the three-way valve to the air and open to the
patient. Ensure that the valve cap is reconnected as the final operation
CO
2
CO
2
measure & standby modes
•
Press the Measure / Standby hard key on the CO
2
/ Anaesthetic Gas module
•
Select the CO
2
numeric area or waveform area to enter the CO
2
/ AG menu
→
Setup
→
Set Operating
Mode to Measure / Standby