These guidelines do 

NOT

 replace the instructions for use and all users should refer to the appropriate 

Operator’s Manual

 for detailed instructions. 

The Illustrations may not necessarily reflect the setup or data displayed on your device. 

 

CAS BeneVision N Series 22/19 & N1 User Guide_V1     Mindray UK © 2020. Errors and omissions excepted. Subject to change. 

Mindray® is a registered trademark of Shenzhen Mindray Bio-Medical Electronics Co. Ltd. 

NIBP 

 

 

Starting & Stopping NIBP measurements 

 

Start and stop NIBP measurement by selecting one of the options  below: 

 

 

 

 

IBP 

 

 

Changing invasive blood pressure labels 

 

Select the IBP numeric area or waveform area to enter the corresponding pressure menu 

 Setup 

 Set IBP1 

or IBP2 Label 

 

Please Note:

 The pressure label is a unique identifier for each type of pressure, therefore, you should select a 

proper pressure label for the source of the pressure you want to  monitor 

 

 

Zeroing pressures 

Zero the transducer by one of the following methods: 

 

Press the Zero hard key on the module 

 

Select the numeric area (such as the Art numeric area), and then select Zero option 

 

Select the IBP Zero quick key 

 

Please Note: 

After the zero calibration is completed, close the three-way valve to the air and open to the 

patient. Ensure that the valve cap is reconnected as the final operation 

 

CO

 

 

CO

measure & standby modes 

 

 

 

 

Press the Measure / Standby hard key on the CO

/ Anaesthetic Gas module 

 

Select the CO

numeric area or waveform area to enter the CO

/ AG menu 

 Setup 

 Set Operating 

Mode to Measure / Standby 