Monitoring ECG/Resp
7-14
7.6.2 Changing the ST Alarm Limits
High and low ST alarm limits can be set individually for each ECG lead. Alarm limits can
also be set separately for single-lead and multi-lead ST monitoring. For 3-lead ECG
monitoring, you can directly change the ST alarm limits in the [
ST Analysis
] menu. For 5- or
12-lead ECG monitoring, you can select [
Adjust ST Limits >>
] from [
ST Analysis
] and then
set the ST alarm limits for each lead.
7.6.3 Adjusting ST Measurement Points
The ST measurement points need to be adjusted if the patient’s heart rate or ECG
morphology changes significantly. To adjust the ST measurement points:
1. In
the
[
ST Analysis
] menu, select [
Adjust ST Points >>
] to open the [
Adjust ST
Points
] window.
2. Select
[
ISO
] or [
ST
] that you need to adjust and then use the and arrow keys to
move the measurement points.
ISO-point (isoelectric): provides baseline for the measurement.
ST: marks the end point of the ST segment analysis.
As shown above, the peak of the R-wave is the reference point for ST measurement. The ST
measurement for each beat complex is the vertical difference between two measurement
points.
NOTE
z
Abnormal QRS complex is not taken into consideration for ST segment analysis.
R-wave peak
P
T
Isoelectric point ST measurement point
Difference=ST value
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...