Batteries
23-2
23.2 Conditioning the Batteries
A battery needs at least two conditioning cycles when it is put into use for the first time. A
battery conditioning cycle is one complete, uninterrupted charge of the battery, followed by
an uninterrupted discharge of the battery. Batteries should be conditioned regularly to
maintain their useful life. Condition the batteries once when they are used or stored for two
months, or when their run time becomes noticeably shorter.
To condition the batteries, follow this procedure:
1. Disconnect the patient monitor from the patient and stop all monitoring and measuring
procedures.
2. Insert the two batteries in need of conditioning into the battery slots of the patient
monitor.
3. Apply AC power to the patient monitor and allow the batteries to charge uninterruptedly
for above 6 hours.
4. Remove AC power and allow the patient monitor to run from the batteries until it shuts
off.
5. Apply AC power again to the patient monitor and allow the batteries to charge
uninterruptedly for above 6 hours.
6. This battery is now conditioned and the patient monitor can be returned to service.
Summary of Contents for BeneView T8
Page 1: ...BeneView T8 Patient Monitor Operation Manual...
Page 2: ......
Page 8: ...2 FOR YOUR NOTES...
Page 24: ...Safety 1 6 FOR YOUR NOTES...
Page 29: ...The Basics 2 5 2 2 4 Rear View Ventilation 1 10 4 3 5 2 9 8 6 7...
Page 74: ...Alarms 6 10 FOR YOUR NOTES...
Page 114: ...Monitoring NIBP 9 10 FOR YOUR NOTES...
Page 124: ...Monitoring IBP 11 8 FOR YOUR NOTES...
Page 142: ...Monitoring Carbon Dioxide 13 12 FOR YOUR NOTES...
Page 172: ...Monitoring RM 17 10 FOR YOUR NOTES...
Page 184: ...Review 19 10 FOR YOUR NOTES...
Page 206: ...Recording 21 8 FOR YOUR NOTES...
Page 210: ...Other Functions 22 4 FOR YOUR NOTES...
Page 256: ...Product Specifications A 24 FOR YOUR NOTES...
Page 290: ...P N 6800 20 50657 1 0...