A - 2
Defibrillator/Monitor Operator’s Manual
A.1.5
Interface Specifications
A.1.6
Signal Outputs Specifications
A.2
Defibrillator Specifications
USB connector
Connects USB flash memory
RJ45 connector
Connects standard network cable.
Multifunctional connector
Connects a cable for analog output or a cable for synchronized
cardioversion.
Multifunctional connector
Standard
Meets the requirements of EN60601-1 for short-circuit protection and
leakage current
ECG Analog Output (only ECG lead set)
Bandwidth
(-3 dB; reference frequency: 10 Hz)
Diagnostic mode: 0.05 to 150 Hz
Monitor mode: 0.5 to 40 Hz
Therapy mode: 1 to 20 Hz
Maximum QRS delay
25 ms (in diagnostic mode, and with Notch off )
Sensitivity
1 V/mV ±5%
Pace enhancement
Signal amplitude: V
oh
≥2.5V
Pulse width: 10ms±5%
Signal rising and falling time: ≤100μs
Synchronous input
Input signal range
0 to 5V (TTL level)
Input impedance
≥10 kΩ
Pulse width
>5 ms
Alarm output
Alarm delay time from the equipment to
other remote equipment
The alarm delay time from the equipment to other remote equipment is ≤4
seconds, measured at the equipment signal output connector.
Alarm signal sound pressure level range
45 db(A) to 85 db(A) within a range of one meter
Standards
Meet standards of IEC 60601-2-4
Defibrillation mode
Manual defib, synchronized cardioversion, AED
Defibrillation waveform
Biphasic truncated exponential (BTE) waveform, auto-compensation
according to patient impedance
Defibrillation electrodes
External paddles set coming with pediatric paddles included, multifunction
electrode pads and internal paddles
Controls and indicators on external
paddles
Charge button, Shock buttons, Energy Select buttons, charge done indicator
Shockable rhythm analysis time
< 8s
Range of selected energy
External defibrillation
1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 100, 150, 170, 200, 300, 360 J
Internal defibrillation
1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50 J
Patient impedance range
External defibrillation
25 to 300 Ω
Summary of Contents for BeneHeart D2
Page 1: ...BeneHeart D3 BeneHeart D2 Defibrillator Monitor Operator s Manual...
Page 2: ......
Page 16: ...10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 32: ...2 12 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 64: ...6 6 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 74: ...7 10 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 86: ...10 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 88: ...11 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 108: ...15 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 110: ...16 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 114: ...18 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 122: ...20 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 150: ...24 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 158: ...25 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 182: ...D 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 192: ...F 4 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 200: ...G 8 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 202: ...H 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 204: ......
Page 205: ...Defibrillator Monitor Operator s Manual I 1 I Declaration of Conformity...
Page 206: ...I 2 Defibrillator Monitor Operator s Manual This page intentionally left blank...
Page 207: ......
Page 208: ...P N 046 010599 00 9 0...