A-5
A.4 Specifications
Parameters
Specifications
Factory Default
Infusion set
standard
Infusion set used in conjunction with
infusion pump should meet the
requirements of
ISO 8536-4:2004
Infusion equipment for medical use
—
Part 4: Infusion sets for single use,
gravity feed, MOD
/
Rate
Unit of Rate (ml/h): 0.1-2000ml/h, the
minimum increment is 0.1ml/h
Unit of Rate (gtt/min): 1-(400*Drip /60)
gtt/min, the minimum increment is
1gtt/min
Note 1: The maximum rate of the
infusion system (infusion pump and
infusion set) is affected by inner and
outer diameter, material, elasticity of
the infusion set and other factors.
Therefore, infusion sets of different
brand and model may differ in
maximum rates.
Note 2: The above declared rates are
based on BOON A2 and B.Braun
Intrafix Safeset infusion sets.
---
Bolus Rate
0.1-2000ml/h
The previous Bolus Rate is
the default value, if not, the
Bolus Rate will be set as the
below rules:
1. If the maximum rate
≥800ml/h, the default bolus
rate is 800ml/h;
2. If the maximum rate
<800ml/h, the default bolus
rate is the maximum rate.
Purge Rate
800ml/h, nonadjustable
800ml/h
VTBI
0.1-9999ml
---
Time
00:00:01-99:59:59 h:m:s
---
Volume
0.1-9999ml
0ml
History Record
Can store up to 1500 history records
/
Select Mode
Rate Mode, Body Weight Mode, Time
Rate Mode
Summary of Contents for BeneFusion VP3 Vet
Page 1: ...BeneFusion VP3 Vet Infusion Pump Operator s Manual...
Page 2: ......
Page 73: ...A 11 Unit of Rate ml h...
Page 90: ...F 1 F Declaration of Conformity P N 046 013101 00 4 0...
Page 91: ......
Page 92: ......