Guidance and manufacturer’s declaration - electromagnetic immunity 

The NeXus-4 is intended for use in the electromagnetic environment specified below. The customer or the user of the 

NeXus-4 should assure that it is used in such an environment. 

Immunity test 

IEC 60601 test level 

Compliance level 

Electromagnetic environment - guidance 

Voltage dips, short 

interruptions and 

voltage variations on 

power supply input 

lines IEC 61000-4-11 

<5 % U

T

 

(>95 % dip in U

T

for 0,5 cycle 

40 % U

T

 

(60 % dip in U

T

for 5 cycles 

70 % U

T

 

(30 % dip in U

T

for 25 cycles 

<5 % U

T

 

(>95 % dip in U

T

for 5 s 

Not  applicable 

Battery powered 

equipment 

Not  applicable 

Battery powered 

equipment 

Not  applicable 

Battery powered 

equipment 

Not  applicable 

Battery powered 

equipment 

 

Power frequency 

(50/60 Hz) magnetic 

field IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic fields should be at 

levels characteristic of a typical location in a 

typical commercial or hospital environment. 

NOTE U

T

 is the a.c. mains voltage prior to application of the test level. 

 

Guidance and manufacturer’s declaration – electromagnetic immunity 

The NeXus-4 is intended for use in the electromagnetic environment specified below. The customer or the user of the 

NeXus-4 should assure that it is used in such an environment. 

Immunity 

test 

IEC 60601 test 

level 

Compliance 

level 

Electromagnetic environment - guidance 

 

 

 

Portable  and  mobile  RF  communications  equipment  should  be 

used no closer to any part of the NeXus-4, including cables, than 

the  recommended  separation  distance  calculated  from  the 

equation applicable to the frequency of the transmitter. 

Recommended separation distance: 

Conducted RF 

IEC 61000-4-6 

3 Vrms 

150 kHz to 80 MHz 

3 V 

1.2√

  

Radiated RF 

IEC 61000-4-3 

3 V/m 

80 MHz to 2.5 GHz 

3 V/m 

1.2√  

  80 MHz to 800 MHz 

2.3√  

  800 MHz to 2.5 GHz 

where 

P

 is the maximum output power rating of the transmitter 

in watts (W) according to the transmitter manufacturer and 

d

 is 

the recommended separation distance in meters (m). 

Field strengths from fixed RF transmitters, as determined by an 

electromagnetic site survey,

a

 should be less than the compliance 

level in each frequency range.

b

 

Interference may occur in the vicinity of equipment marked with 

the following symbol:  

 

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. 

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection 

from structures, objects, and people. 

a

 

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile 

Summary of Contents for NeXus-4

Page 1: ...NeXus 4 User Manual Version 2 0 0344...

Page 2: ...lay operators 7 3 Product overview 8 3 1 Product components 8 3 2 Intended use 8 3 3 NeXus 4 views 9 3 4 User interface 9 3 5 Patient connections 10 3 6 Device label 10 4 Instructions for use 11 4 1 S...

Page 3: ...RMA number in case a return is required Never ship products back to Mind Media without this authorization and or RMA number 1 2 Warranty information The product except its cables and accessories is w...

Page 4: ...tion read important safety information Important information guidance for use Consult instructions for use Device has type CF applied parts Ingress protection rating Keep dry CE certified 93 42 EC Ann...

Page 5: ...esonance Imaging MRI equipment or equipment capable of emitting diagnostic levels of ionizing radiation Apart from the above there are no contra indications There are no known side effects from the us...

Page 6: ...used in conjunction with accessories that are not manufactured or approved by the manufacturer No modification of this product is allowed The product should not be tampered with Do not touch the elect...

Page 7: ...PC receiver and the device the signal transmission is not as reliable Do not use an operating cellular phone within 50 cm of the device to avoid excessive noise on the signals Sharp bends or winding...

Page 8: ...bsite www mindmedia com 3 2 Intended use The product is intended to be used for acquisition of electro physiological signals by or under supervision of a physician The user must have knowledge of curr...

Page 9: ...ill result in a signal in the digital channel of the product In this case the device will not shut down Status LED The table below states all possible states of the LED indicators of the device LED In...

Page 10: ...ments Leads that fit in the bipolar inputs have two bipolar cables with 2 snaps per cable going to the patient The bipolar input uses a 6 pin connector Auxiliary Used for sensors that require 5V power...

Page 11: ...B HDD 50 GB 250 GB recommended Internet connection or CD DVD Drive Operating system Windows Windows 10 64 bit Windows 8 1 64 bit Windows 7 32 bit 64 bit Important Disconnect all Mind Media products fr...

Page 12: ...When the device is not in use e g there is no active connection pairing with the PC the NeXus 4 will switch off after 5 minutes to save its power 4 3 Transfer data to PC Bluetooth IT network connectio...

Page 13: ...til Windows reports that the Device is ready to use The Bluetooth pairing remains valid until you plug the Bluetooth dongle into a different USB port delete the link from Windows or pair the device wi...

Page 14: ...o high pass or low pass filters that can cause signal phase shifts or filter overflows are present The NeXus 4 has one dual bipolar input This connector leads to two bipolar input stages for in total...

Page 15: ...submerge product in liquids Never use sharp tools or aggressive chemicals for cleaning or disinfecting Do not sterilize equipment Environmental protection Special EU instructions for disposal are appl...

Page 16: ...age power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Battery powered equipment Voltage fluctuations flicker emissions 61000 3 3 N...

Page 17: ...or the user of the NeXus 4 should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Portable and mobile RF communic...

Page 18: ...NeXus 4 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the NeXus 4 can help prevent electromagnetic interference by...

Page 19: ...Class CF Applied parts Its enclosure also after removal of battery cover including all contacts and receptacles The patient accessories Accessible parts Apart from the applied parts the NeXus 4 has no...

Page 20: ...ry Batteries 2 x AA disposable alkaline 1 5V Power Saving 5 minutes without connection to PC Battery low indication level 2 0 V 0 1V Battery empty shut down level 1 7 V 0 1V Bluetooth Communication Bl...

Page 21: ...Digital channel bits 1 bit 0 to 255 0x01 1 ON OFF button 0x02 Battery state 0 battery empty 63 battery full alkaline 0x04 0x08 0x10 0x20 0x40 0x80 1 battery empty 6 Saw Sawtooth test signal 256 sample...

Page 22: ...22 Copyright 2017 Mind Media All rights reserved www mindmedia com...

Reviews: