MindMedia NeXus-4 User Manual Download Page 14

 

14 

 

5

 

Operational principles 

5.1

 

Bipolar input channels 

The  input  stage  for  measuring  bipolar  electrophysiological  signals  is  configured  as  an 

instrumentation  amplifier.  The  difference  between  a  ‘plus’  and  ‘minus’  signal  is  amplified.  The 

patient ground electrode is required to keep patient potential and amplifier potential at about the 

same level. 

All electrode cables are shielded with the average of the ‘plus’ and ‘minus’ electrode signal (active 

shielding).  The  active  shielding  ensures  that  disturbances  such  as  cable  movement  artifacts  and 

mains interference (50/60 Hz) are reduced to a minimum. 

After the first amplifier stage (gain = 20) the signals go directly to the ADC. No high pass or low pass 

filters that can cause signal phase shifts or filter overflows are present. 

The NeXus-4 has one dual bipolar input. This connector leads to two bipolar input stages for in total 

two differential channels.  

5.2

 

Auxiliary input channel 

The  Auxiliary  input  has  a  5-pin  connector.  Signals  on  this  connector  are  +5V  output,  -5V  output, 

GND, +signal input and -signal input. The +5V/-5V/GND pins can be used to power an active sensor. 

The + and - inputs are connected to an instrumentation amplifier with a gain of 1. The output of the 

amplifier goes to the ADCs without any filtering. 

5.3

 

Filtering 

The  ADC  of  the  device  has  a  digital  sinc  filter  with  a  cutoff  frequency  of  0.2  *  sample  frequency. 

Supported sampling rates are defined in 

chapter 8

. There is a 1

st

 order low pass filter before the ADC 

with a -3db point at 408Hz.  

Besides above no other filtering is applied. 

 

Summary of Contents for NeXus-4

Page 1: ...NeXus 4 User Manual Version 2 0 0344...

Page 2: ...lay operators 7 3 Product overview 8 3 1 Product components 8 3 2 Intended use 8 3 3 NeXus 4 views 9 3 4 User interface 9 3 5 Patient connections 10 3 6 Device label 10 4 Instructions for use 11 4 1 S...

Page 3: ...RMA number in case a return is required Never ship products back to Mind Media without this authorization and or RMA number 1 2 Warranty information The product except its cables and accessories is w...

Page 4: ...tion read important safety information Important information guidance for use Consult instructions for use Device has type CF applied parts Ingress protection rating Keep dry CE certified 93 42 EC Ann...

Page 5: ...esonance Imaging MRI equipment or equipment capable of emitting diagnostic levels of ionizing radiation Apart from the above there are no contra indications There are no known side effects from the us...

Page 6: ...used in conjunction with accessories that are not manufactured or approved by the manufacturer No modification of this product is allowed The product should not be tampered with Do not touch the elect...

Page 7: ...PC receiver and the device the signal transmission is not as reliable Do not use an operating cellular phone within 50 cm of the device to avoid excessive noise on the signals Sharp bends or winding...

Page 8: ...bsite www mindmedia com 3 2 Intended use The product is intended to be used for acquisition of electro physiological signals by or under supervision of a physician The user must have knowledge of curr...

Page 9: ...ill result in a signal in the digital channel of the product In this case the device will not shut down Status LED The table below states all possible states of the LED indicators of the device LED In...

Page 10: ...ments Leads that fit in the bipolar inputs have two bipolar cables with 2 snaps per cable going to the patient The bipolar input uses a 6 pin connector Auxiliary Used for sensors that require 5V power...

Page 11: ...B HDD 50 GB 250 GB recommended Internet connection or CD DVD Drive Operating system Windows Windows 10 64 bit Windows 8 1 64 bit Windows 7 32 bit 64 bit Important Disconnect all Mind Media products fr...

Page 12: ...When the device is not in use e g there is no active connection pairing with the PC the NeXus 4 will switch off after 5 minutes to save its power 4 3 Transfer data to PC Bluetooth IT network connectio...

Page 13: ...til Windows reports that the Device is ready to use The Bluetooth pairing remains valid until you plug the Bluetooth dongle into a different USB port delete the link from Windows or pair the device wi...

Page 14: ...o high pass or low pass filters that can cause signal phase shifts or filter overflows are present The NeXus 4 has one dual bipolar input This connector leads to two bipolar input stages for in total...

Page 15: ...submerge product in liquids Never use sharp tools or aggressive chemicals for cleaning or disinfecting Do not sterilize equipment Environmental protection Special EU instructions for disposal are appl...

Page 16: ...age power supply network that supplies buildings used for domestic purposes Harmonic emissions IEC 61000 3 2 Not applicable Battery powered equipment Voltage fluctuations flicker emissions 61000 3 3 N...

Page 17: ...or the user of the NeXus 4 should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Portable and mobile RF communic...

Page 18: ...NeXus 4 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the NeXus 4 can help prevent electromagnetic interference by...

Page 19: ...Class CF Applied parts Its enclosure also after removal of battery cover including all contacts and receptacles The patient accessories Accessible parts Apart from the applied parts the NeXus 4 has no...

Page 20: ...ry Batteries 2 x AA disposable alkaline 1 5V Power Saving 5 minutes without connection to PC Battery low indication level 2 0 V 0 1V Battery empty shut down level 1 7 V 0 1V Bluetooth Communication Bl...

Page 21: ...Digital channel bits 1 bit 0 to 255 0x01 1 ON OFF button 0x02 Battery state 0 battery empty 63 battery full alkaline 0x04 0x08 0x10 0x20 0x40 0x80 1 battery empty 6 Saw Sawtooth test signal 256 sample...

Page 22: ...22 Copyright 2017 Mind Media All rights reserved www mindmedia com...

Reviews: