background image

 

[4] 

Red pin

 is anodal (+), 

which enhances neuronal 
activity.  

Black pin

 is cathodal (-), 

which inhibits neuronal 
activity.

 

USING tDCS 

IMPORTANT 

If at any time the client has an adverse reaction, disconnect one or both 
electrodes or unplug the cable to pause the session. You may also press 
the power button   to end the session. Unlike CES, the intensity cannot 
be manually adjusted and preferences cannot be set for tDCS sessions. 

ELECTRODE PLACEMENT  AND SETUP 

In order to perform tDCS, you must always connect two electrodes to the 
body in order to have a complete circuit. Your tDCS kit includes a high 
quality cable with pin connectors. There are also four carbon electrodes: 
two small, one large and one long thin. Place a small electrode over the 
specific site to be treated. Use long thin electrode for treating two sites 
simultaneously such as FP1 and FP2 (for ADHD) or FZ and CZ, then CZ 
and PZ (for OCD). Use the large electrode 
as the receiver to complete the circuit. 
Soak the sponges in tap water or an 
appropriate saline solution, and place the 
small sponge on the area of the head 
being treated. Put the small electrode 
inside the sponge pouch. Place the large 
electrode inside the large sponge pouch 
on the opposite side of the body (eg. 
opposite side of the head, face or 
opposite shoulder). Use the headband to secure the sponges and 
electrodes to the head. 

Appropriate saline solutions include contact lens saline solution, or a 
mixture of salt and tap water (½tsp / 500ml).

 

 

Summary of Contents for Oasis Pro

Page 1: ...TRANSCRANIAL DC STIMULATION PROTOCOL GUIDE For use with the Oasis Pro...

Page 2: ...ility of the clinician to determine proper electrode placement and appropriate and safe current usage Mind Alive Inc does not claim any medical benefits from the use of tDCS nor claim to have any expe...

Page 3: ...generated by putting a wet cloth cotton ball or sponge on the scalp over the brain regions to be affected and or a sponge or cotton ball on the shoulder or the face The tDCS sessions deliver a maximum...

Page 4: ...ession There are six 6 tDCS sessions 0 25mA 0 50mA 0 75mA 1 00mA 1 50mA and 2 00mA Each of these sessions is 20 minutes long Additionally two slots are available to store custom tDCS sessions 1 and 2...

Page 5: ...or the CES sessions The graph will display one of the following codes to indicate the status of the connection The tDCS session will begin automatically only when a good connection is detected Open Ba...

Page 6: ...e are also four carbon electrodes two small one large and one long thin Place a small electrode over the specific site to be treated Use long thin electrode for treating two sites simultaneously such...

Page 7: ...read using a computer When used with a computer both DAVID Session Editor files and audio WAV files are available Download the sessions to the Oasis Pro using the audio files or a CD player Connect t...

Page 8: ...800 661 MIND 6463 780 465 MIND 6463 info mindalive com www mindalive com 6716 75 St NW Edmonton Alberta Canada T6E 6T9 Mind Alive Inc...