background image

10

Specifications

Input:

  Type E, J, K, or T thermocouples.

Output:

  4 to 20 mA DC over specified range.

Accuracy:

   TT190:  +/- 0.2% of span.
   TT230:  +/- 0.3% of span when factory calibrated.

Linearity:

  Voltage linear.

Ambient Temperature:
   

Operating:  -40 to 85°C (-40 to 185°F ).

   Storage:  -55 to  100°C (-67 to 212°F ).

Ambient Temperature Effects:

   TT190: +/- 0.018% of span°C.
   TT230: +/- 0.025% of span°C.

Cold Junction Compensation:
   

+/- 0.03°C/°C for -25 to 70°C ambients.

   +/- 0.06°C/°C for -40 to -25 and 70 to 85°C ambients.

Adjustments:

   TT190:  Zero and Span, +/- 5% of span.
   TT230:  Depends on temperature range selected.

Warmup Drift:

  +/- 0.2% of span max., assuming  Vsupply = 24

    VDC and Rloop = 250 ohms.  Stable     within 30 minutes.

Supply Voltage:

  10 to 35 volts DC with no load.

   Reverse polarity protected.

Voltage effect:

  +/- 0.001% of span per volt.

Maximum Load Resistance:

  The maximum allowable

   resistance of the signal-carrying loop is given by this
   formula:  Rloop max = (Vsupply-10)/.02 amps.

Maximum Output Current:

  27 mA.

Connections:

 Terminal blocks for wires from AWG 22 to

   AWG 14.

Physical:

  Epoxy potted for moisture resistance.

Weight:

  1.8 oz. (58 grams).

Model Number Coding:

TT190    JU  1   K    . . . Sample part number

                              

Temperature Range:

  Use one of the over 175 range codes in the MINCO Temptran sales

bulletin.  Note: A range that is available for a particular Temptran model is not automatically
available for the TT190 or TT230; there may be a nominal charge.  Also, ranges available for the
TT190 and TT230 depend on T/C type.

Output:

  1 = 4 to 20 mA DC.

                   

Calibrated to match T/C Type:

                    

EU  = Type E  ( Chromel - Constantan ), Ungrounded Junction

                   

JU  = Type J  ( Iron - Constantan ), Ungrounded Junction

                   

KU = Type K ( Chromel - Alumel ), Ungrounded Junction

                   

TU = Type T ( Copper - Constantan ), Ungrounded Junction

                      

Model Number:

  TT190 T/C Temptran

                      TT230 Rangeable T/C Temptran

When quality and performance are as important as price, call...

MINCO

MINCO

 PRODUCTS, INC.

7300 Commerce Lane/Minneapolis, Minnesota 55432-3177 U.S.A.

Telephone:(763)571-3121                     FAX:(763)571-0927
©

2001 MINCO PRODUCTS, INC.

Summary of Contents for TT190

Page 1: ...emperature Transmitters Installation and Operating Instructions INTRINSICALLY SAFE NON INCENDIVE MINCO MINCO PRODUCTS INC 7300 Commerce Lane Minneapolis Minnesota 55432 3177 U S A Telephone 763 571 31...

Page 2: ...re to installation requirements as set forth by the National Electrical Code NEC and any other applicable codes and standards See pages 3 7 for further information on installing these transmitters in...

Page 3: ...ure requirements may be found in ANSI ISA Standards S82 01 S82 02 and S82 03 1 1 Accessibility The transmitters must be installed within the enclosure so that its circuits are accessible by the use of...

Page 4: ...nch in diameter the end of which is rounded The force shall be applied for one minute to any point on the overall enclosure except the bottom The bottom shall sustain a force of 65 Newtons 15 pounds c...

Page 5: ...5...

Page 6: ...6...

Page 7: ...7...

Page 8: ...hen the span 200 50 250 C Temperature Span T C Type K E J K T K Connect RB range RC range RE range 1 A 50 to 70 C 150 to 200 C 300 to 400 C 2 A 70 to 100 C 200 to 300 C 400 to 600 C 3 A 100 to 200 C 3...

Page 9: ...epeated However it is recommended that step 5 be repeated to check that meter still reads 4 mA to ensure proper calibration Warranty Items returned within one year from the date of sale transportation...

Page 10: ...ula Rloop max Vsupply 10 02 amps Maximum Output Current 27 mA Connections Terminal blocks for wires from AWG 22 to AWG 14 Physical Epoxy potted for moisture resistance Weight 1 8 oz 58 grams Model Num...