TROUBLESHOOTING GUIDE: 

 

Condition 

Causes 

Remedy 

Display shows "1" 
continuously in only the 
leftmost digit. 

RTD is open. 

Disconnect from meter, 
ohm RTD continuity and 
connections. Replace the 
RTD if faulty. 

 

Over range temperature. 

Switch to °F scale. 
System is only accurate to 
400°F, but meter will read 
to 600°F. 

Reading shows "-1" 
continuously in only the 
leftmost digit. 

The RTD is shorted. 

Disconnect from meter, 
ohm RTD continuity and 
connections. Replace the 
RTD if faulty. 

" " or "LO BAT" reading in 
upper left corner of LCD 
display. 

Battery is worn out. 

Replace battery. 

 

Poor battery connections. 

Clean connections. 

LCD display shows 
nothing. 

Dead or missing battery. 

Replace battery. 

Display shows very unlikely 
temperature. 

Wrong scale setting. 

Move switch on front panel 
to appropriate position. 
(Fahrenheit or Celsius). 

 

Faulty RTD. 

Ohm the RTD connections. 
Replace if faulty. 

 

Wrong Element type. (e.g. 
The PD in S902

PD

60 must 

match the meter's code 
TI142

PD

.) 

Use correct RTD. Meter's 
code is listed inside battery 
compartment. 

 