7

 

 

4.1.6 

You can skip any calibration points by pressing “ALARM” button until the unit will exit 

 

 

   

and turn off.

 

 

4.1.7 

Once exit, the unit will default to Normal temperature Mode.

 

 4.2 

KTC Software Calibration Check using Temperature Calibrator (Fluke 714 or 5500A/5520A)

 

 

4.2.1 

Condition the unit to be calibrated in a stable environment as described in 3.1 to ensure 

 

 

   

they are in thermal equilibrium.

 

 

4.2.2 

Set the Fluke714 or 5500A/5520A to be 0°C and “K” type mode.

 

 

4.2.3 

Connect the KTC cable to Fluke714 or 5500A/5520A. Press and hold the “ALARM” 

 

 

   

button, then apply power. Press the “MEASURE” button 4 times. After 5 seconds, the 

 

 

   

unit will enter into KTC calibration mode.

 

 

4.2.4 

Wait 60s for first point to stable and Press the “MEASURE” button to confirm. The next 

 

 

   

calibration point 50°C will appear.

 

 

4.2.5 

Set the Fluke714 or 5500A/5520A to 50°C. Wait for 30 seconds and then press the 

 

 

   

“MEASURE” button to confirm. (For retest, the stable time is 10s).

 

 

4.2.6 

Repeat steps 4.2.4~4.2.5 to complete all calibration points per table below.

 

Calibration points 

Stable time for calibration  Stable time for retest 

Fluke

 

0      (Thermocouple) 

60s 

60s 

714 or 5500A/5520A

 

50     (Thermocouple) 

30s 

10s 

714 or 5500A/5520A

 

100    (Thermocouple) 

30s 

10s 

714 or 5500A/5520A

 

250    (Thermocouple) 

30s 

10s 

714 or 5500A/5520A

 

450    (Thermocouple) 

30s 

10s 

714 or 5500A/5520A

 

-20    (Thermocouple) 

30s 

10s 

714 or 5500A/5520A

  

 

 

4.2.7 

You can skip any calibration 

 

 

   

points by pressing “ALM” button 

 

 

   

until the unit will exit and 

 

 

   

turn off.

 

 

4.2.8 

Once exit, the unit will default to 

 

 

   

Normal Temperature Mode.

                       

Summary of Contents for TEMP-GUN 2266-20

Page 1: ...TRUCTIONS Calibration MILWAUKEE TOOL 13135 WEST LISBON ROAD BROOKFIELD WISCONSIN 53005 2550 USA 58 92 2266d3 Calibration and verification are performed at an ambient temperature of 23 C 25 C and ambie...

Page 2: ...Specifications 3 Basic Maintenance 4 Loading Changing the Batteries 4 Maintaining Tool 4 Cleaning the Laser Windows 4 Scope 5 Equipment Use 5 Testing Condition and Interference 5 Field calibration Pr...

Page 3: ...nual If you do not do so the protection provided by the Laser TEMP Gun may be impaired Read the Safety Information page before servicing this product In this manual a WARNING identifies conditions and...

Page 4: ...ctices that could result in personal injury or loss of life Specific precautions To reduce the risk of injury user must read and understand operator s manual Laser light Do not stare into beam or view...

Page 5: ...2 C 4 F or 2 of reading whichever is greater Assume ambient operating temperature of 23 C to 25 C 73 F to 77 F Min measuring distance 2 50 C 122 F 4 50 C 122 F Display resolution 0 1 C 0 1 F Contact...

Page 6: ...r operate at full power with fully charged batteries clean the contacts on the battery door If the tool still does not work properly return the tool to a MILWAUKEE service facility for repairs Cleanin...

Page 7: ...od 3 0 TESTING CONDITION AND INTERFERENCE 3 1 Calibration and verification are performed at an ambient temperature of 23 C 25 C and ambient relative humidity 55 in a controlled room Each thermometer s...

Page 8: ...ter 5 seconds the unit will enter into IR calibration mode 4 1 3 Press the ALARM button to skip 25 C and move to 0 C calibration point 4 1 4 Put the unit on the test fixture Wait for 3 4 seconds and t...

Page 9: ...2 4 Wait 60s for first point to stable and Press the MEASURE button to confirm The next calibration point 50 C will appear 4 2 5 Set the Fluke714 or 5500A 5520A to 50 C Wait for 30 seconds and then pr...

Reviews: