MILWAUKEE ELECTRIC TOOL CORPORATION – TEST & MEASUREMENT PRODUCT 

 

REPAIR SERVICE INSTRUCTIONS — CALIBRATION 

2266-20 Laser TEMP-GUN™ Thermometer 

 

Environmental 

Condition 

 

 
 

REPAIR SERVICE 

INSTRUCTIONS 

Calibration 

 

 

 

 

 

 

 

 

 

 

 

 

 

MILWAUKEE TOOL

 

 

 

13135 WEST LISBON ROAD 

 BROOKFIELD, WISCONSIN 53005-2550 

 USA 

58-92-2266d3

Calibration and verification are performed at an ambient temperature of 23°C ~ 25°C 

and ambient relative humidity <55% in a controlled room. Each thermometer should 

be uniquely identified as to the room and/or area in which it is used or tested.

www.milwaukeetool.com

03/2014

Summary of Contents for TEMP-GUN 2266-20

Page 1: ...TRUCTIONS Calibration MILWAUKEE TOOL 13135 WEST LISBON ROAD BROOKFIELD WISCONSIN 53005 2550 USA 58 92 2266d3 Calibration and verification are performed at an ambient temperature of 23 C 25 C and ambie...

Page 2: ...Specifications 3 Basic Maintenance 4 Loading Changing the Batteries 4 Maintaining Tool 4 Cleaning the Laser Windows 4 Scope 5 Equipment Use 5 Testing Condition and Interference 5 Field calibration Pr...

Page 3: ...nual If you do not do so the protection provided by the Laser TEMP Gun may be impaired Read the Safety Information page before servicing this product In this manual a WARNING identifies conditions and...

Page 4: ...ctices that could result in personal injury or loss of life Specific precautions To reduce the risk of injury user must read and understand operator s manual Laser light Do not stare into beam or view...

Page 5: ...2 C 4 F or 2 of reading whichever is greater Assume ambient operating temperature of 23 C to 25 C 73 F to 77 F Min measuring distance 2 50 C 122 F 4 50 C 122 F Display resolution 0 1 C 0 1 F Contact...

Page 6: ...r operate at full power with fully charged batteries clean the contacts on the battery door If the tool still does not work properly return the tool to a MILWAUKEE service facility for repairs Cleanin...

Page 7: ...od 3 0 TESTING CONDITION AND INTERFERENCE 3 1 Calibration and verification are performed at an ambient temperature of 23 C 25 C and ambient relative humidity 55 in a controlled room Each thermometer s...

Page 8: ...ter 5 seconds the unit will enter into IR calibration mode 4 1 3 Press the ALARM button to skip 25 C and move to 0 C calibration point 4 1 4 Put the unit on the test fixture Wait for 3 4 seconds and t...

Page 9: ...2 4 Wait 60s for first point to stable and Press the MEASURE button to confirm The next calibration point 50 C will appear 4 2 5 Set the Fluke714 or 5500A 5520A to 50 C Wait for 30 seconds and then pr...