SPECIFICATIONS 7
Calibration
EC / TDS
Single cell factor calibration
6 standards:
84 µS/cm, 1413 µS/cm, 5.00 mS/cm,
12.88 mS/cm, 80.0 mS/cm, 111.8 mS/cm
One-point offset: 0.00 µS/cm
Salinity
One-point
with MA9066 Salinity calibration solution
Temp.
No temperature calibration
Temperature
compensation
ATC – automatic
MTC – manual, without temperature probe
–20.0 to 120.0 °C (–4.0 to 248.0 °F)
NO TC – without temperature compensation
Conductivity
temp. coefficient
0.00 to 6.00 % / °C (EC & TDS only)
Default value: 1.90 % / °C
TDS factor
0.40 to 0.80
Default value: 0.50
Logging memory
Max. 1000 log records (stored in up to 100 lots)
On demand, 200 logs
On stability, 200 logs
Interval logging, 1000 logs
PC connectivity
1 micro USB port
Battery type
3 x 1.5V alkaline AA (included)
Battery life
Approx. 200 hours of use
Environment
0 to 50 °C; maximum RH 95%
Dimensions
200 x 85 x 50 mm (7.9 x 3.3 x 2.0”)
Casing
IP67 protection level
Weight
260 g (0.57 lb)
PROBE SPECIFICATIONS
EC probe
MA815D/1
Temperature range 0 to 60 °C (32 to 140 °F)
Temperature sensor NTC10K
4-ring type
Stainless steel
Connector socket
DIN, 7 pins
Body
ABS
Dimensions
total length: 198 mm (7.8")
active part: 96 mm (3.8")
Ø 16 mm (0.63")
Cable length
1 m (3.2 ft)