LOGGING 37
Interval Logging
1. From the Setup mode, set Log Type to INTERVAL (default) and
desired time interval.
2. From the measurement screen press LOG/CLEAR.
LCD displays “PLEASE WAIT”. The LOG ### LOT ### screen
displays on third LCD line the measurement log number
(bottom left) and interval logging session lot number (bottom
right).
3. Press RANGE/
during logging to display the number of
available records ("FREE" ###). Press RANGE/
again to
return to return to active logging screen.
4. Press LOG/CLEAR again (or ESC) to end current interval
logging session.
LCD displays “LOG STOPPED”. Meter returns to measurement
screen.
Interval Logging Warnings
“OUT OF SPEC” Sensor failure is detected. Loggings stops.
“MAX LOTS”
Maximum number of lots reached (100).
Cannot create new lots.
“LOG FULL”
Log space is full (1000 logs limit was reached).
Loggings stops.