background image

MEASUREMENTS 

7

6. MEASUREMENTS

Make sure the meter has been calibrated and the 

protective cap has been removed.
Immerse the tip of the probe in the sample to be 

tested while ensuring that also the temperature 

sensor is immersed.
For accurate dissolved oxygen measurements, a water movement 

of at least 0.3 m/sec is required. This is to ensure that the 

oxygen-depleted membrane surface is constantly replenished. 

A moving stream will provide an adequate circulation. To quickly 

check if the water speed is sufficient, wait for the reading to 

stabilize and then move the D.O. probe. If the reading remains 

stable, the measurement conditions are right, while if the reading 

increases the water movement is not adequate.
During field measurements, this condition may be met by manually 

agitating the probe. Accurate readings are not possible while the 

liquid is at rest.
During laboratory measurements, the use of a magnetic stirrer to 

ensure a certain velocity in the fluid is recommended.
In this way, errors due to the diffusion of the oxygen present in the 

air in the solution are reduced to a minimum.
Anyway, wait the time necessary for thermal equilibrium to occur 

between the probe and the sample (a few minutes for temperature 

difference of several degrees).
The meter can display D.O. readings in mg/L (ppm) and in % O

(rate of Oxygen saturation with reference to 100% at sea level). 

Press the RANGE key to select the desired range.
The temperature measurement is always displayed on the lower 

part of the LCD.

 

If the sample contains salts or if the measurement is performed 

at altitude different from sea level, the readout values must be 

corrected, taking into account the lower degree of oxygen solubility.
Set altitude and salinity compensations before performing calibration 

and taking measurements. The meter will automatically compensate 

for these factors.

Summary of Contents for MI605

Page 1: ...INSTRUCTION MANUAL SATISFACTION GUARANTEED MI605 Dissolved Oxygen Meter...

Page 2: ...is prohibited without the written consent of the copyright owner Milwaukee Instruments Inc Rocky Mount NC 27804 USA THANK YOU for choosing Milwaukee Instruments This instruction manual will provide yo...

Page 3: ...AL OPERATIONS 4 4 PROBE PREPARATION 5 5 PROBE POLARIZATION 6 6 MEASUREMENTS 7 7 ALTITUDE COMPENSATION 8 8 SALINITY COMPENSATION 9 9 CALIBRATION PROCEDURE 10 10 PROBE MEMBRANE MAINTENANCE 11 11 BATTERY...

Page 4: ...m resolution SALINITY COMPENSATION 0 to 80 g L 1 g L resolution PROBE MA840 included ENVIRONMENT 0 to 50 C 100 RH max BATTERY TYPE 1 x 9V alkaline included BATTERY LIFE approx 100 hours of use AUTO O...

Page 5: ...ied with the meter with electrolyte while shaking it gently Refill with clean electrolyte 4 Gently tap the sides of the membrane with the finger tip to ensure that no air bubbles remain trapped To avo...

Page 6: ...consumed by passing through the sensitive diaphragm and dissolving in the electrolyte solution inside the probe If this operation is interrupted the electrolyte solution continues to be enriched with...

Page 7: ...e not possible while the liquid is at rest During laboratory measurements the use of a magnetic stirrer to ensure a certain velocity in the fluid is recommended In this way errors due to the diffusion...

Page 8: ...7 12 2 11 9 11 4 11 0 10 6 39 2 6 12 4 12 0 11 6 11 2 10 8 10 4 10 1 42 8 8 11 8 11 4 11 0 10 6 10 3 9 9 9 6 46 4 10 11 3 10 9 10 5 10 2 9 8 9 5 9 2 50 0 12 10 8 10 4 10 1 9 7 9 4 9 1 8 8 53 6 14 10...

Page 9: ...matically compensate for salinity effects see also following table for verification C Salinity g l at Sea Level F 0 g l 10 g l 20 g l 30 g l 35 g l 10 11 3 10 6 9 9 9 3 9 0 50 0 12 10 8 10 1 9 5 8 9 8...

Page 10: ...nditioning mode automatic polarization y Once COND disappears the probe is polarized and the instrument can be calibrated y Remove the probe protective cap y Press CAL appears on the larger LCD togeth...

Page 11: ...ntly tap the sides of the membrane with the finger tip to ensure that no air bubbles remain trapped Do no directly tap the bottom with the finger as this will damage the membrane y Makesurethattherubb...

Page 12: ...ts from the environment but allows oxygen to enter If any dirt is observed on the membrane rinse it carefully with distilled or deionized water If any imperfections still exist or any damage is eviden...

Page 13: ...y life remain A low battery will result in unreliable measurements It is recommended to replace the battery immediately by using an alkaline 9V battery Turn the meter off unscrew the two screws on the...

Page 14: ...duct makesureitisentirelysuitableforyourspecific application and for the environment in which it is used Any modification introduced by the user to the supplied equipment may compromise the meter s pe...

Page 15: ...MilwaukeeInstrumentsreservestherighttomakeimprovementsindesign construction and appearance of its products without advance notice...

Page 16: ...C H 6726 Szeged HUNGARY tel 36 62 428 050 fax 36 62 428 051 www milwaukeeinst com e mail sales milwaukeeinst com Milwaukee Instruments Inc 2950 Business Park Drive Rocky Mount NC 27804 USA tel 1 252 4...

Reviews: