MA9011
3.5M KCl Electrolyte Solution. For faster response, unscrew the fill hole screw
during measurements.
STORAGE PROCEDURE
To minimize clogging and assure a quick response time, the glass bulb and the junction of
the electrode should be kept moist and not allowed to dry out.
Replace the solution in the protective cap with a few drops of
MA9015
Storage Solution
or, in its absence, Filling Solution (
MA9011
). Follow the Preparation Procedure before
taking measurements.
Note: NEVER STORE THE ELECTRODE IN DISTILLED OR DEIONIZED WATER.
PERIODIC MAINTENANCE
Inspect the electrode and the cable. The cable used for connection to the instrument must
be intact and there must be no points of broken insulation on the cable or cracks on the
electrode stem or bulb. Connectors must be perfectly clean and dry. If any scratches or
cracks are present, replace the electrode. Rinse off any salt deposits with water.
Probe Maintenance
Refill the reference chamber with fresh electrolyte (
MA9011
). Allow the electrode to
stand upright for 1 hour. Follow the Storage Procedure above.
CLEANING PROCEDURE
•
Wine deposits
Soak in Milwaukee
MA9016
cleaning solution for 1 hour
•
Wine stains
Soak in Milwaukee
MA9016
cleaning solution for 1 hour
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thor-
oughly with distilled water, refill the reference chamber with fresh electrolyte and soak the
electrode in
MA9015
Storage Solution for at least 1 hour before taking measurements.