24
Instruction Manual Mi170 Bench Meter
Press the arrow
keys to change the displayed value.
Press the GLP/ACCEPT key to accept the value or the CAL key to escape.
Press the arrow
keys to select the next/previous parameter.
Press the SETUP key to exit SETUP menu at any time.
The following table lists the SETUP parameters, their valid values range and the fac-
tory settings (default):
Item
Description
Valid values
Default
CELL
Cell Constant
0.500 to 1.700
1.000
tdS
TDS Factor
0.40 to 0.80
0.50
tc
Temperature Coefficient
0.00 to 6.00%/
°
C
1.90
rEF
Reference Temperature
20.0 or 25.0
°
C
25.0
TIME
Time (hh:mm)
00.00 to 23.59
Current
DATE
Date (MM.DD.YYYY)
01.01.2000-12.31.2099
Current
bEEP
Beep Status
ON/OFF
OFF
bAud
Baud Rate
600; 1200; 2400; 4800; 9600
2400
In Id
Instrument ID
0000 to 9999
0000
tEMP
Temperature Unit
°
C or
°
F
°
C
9
www.milwaukeeinst.com
Notes:
• If the reading is out of range, the full-scale value (100.0 for
MTC
/
ATC
mode or
400.0 for uncompensated TDS) will be displayed blinking.
• If the SHIFT&RANGE/FIXED keys are pressed to freeze the LCD range and the reading
goes out of range, the full-scale value of the frozen range will be displayed blink-
ing.
NaCl MEASUREMENTS
• Press the RANGE/FIXED key to enter NaCl measurement mode.
• The instrument will display the NaCl reading on the primary LCD and the temperature
on the secondary LCD, along with the reference temperature.
AUTORANGING
The EC and TDS scales are autoranging. The meter automatically sets the scale with
the highest possible resolution.
By pressing the SHIFT&RANGE/FIXED keys, the autoranging feature is disabled and the
current range is frozen on the LCD. The “Auto” “OFF” (autoranging disabled) message
will be displayed on the LCD for a few seconds. To restore the autoranging option,
press the SHIFT&RANGE/FIXED keys again. The “Auto” “On” (autoranging enabled)
message will be displayed on the LCD for a few seconds.
Note
: Autoranging is automatically restored if the range is changed, if the setup or
calibration modes are entered and if the meter is turned off and back on again.