30
Instruction Manual Mi170 Bench Meter
ACCESSORIES
MA9060
12880 µS/cm Calibration Solution (230 ml bottle)
MA9061
1413 µS/cm Calibration Solution (230 ml bottle)
MA9063
84 µS/cm Calibration Solution (230 ml bottle)
MA9065
111.8 mS/cm Calibration Solution (230 ml bottle)
MA9069
5000 µS/cm Calibration Solution (230 ml bottle)
MA9064
80000 µS/cm Calibration Solution (230 ml bottle)
MA9066
NaCl 100% Calibration Solution (230 ml bottle)
MA9310
12 VDC Adapter, 220 V
MA9311
12 VDC Adapter, 110 V
MA9315
Electrode Holder
MA814DB/1
EC/Temperature Probe
MA9350
RS232 connection cable (2 meters)
TROUBLESHOOTING
SYMPTOMS
PROBLEM
SOLUTION
Reading fluctuates up and
down (noise).
EC probe sleeve not properly
inserted; air bubbles inside
sleeve.
Insert the sleeve. Tap the
probe to remove air bubbles.
Display shows EC, TDS or
NaCl reading blinking.
Reading out of range.
Recalibrate the meter;
Check the sample is within
measurable range;
Make sure the autoranging
feature is not enabled.
Meter fails to calibrate or
gives faulty readings.
Broken EC probe.
Replace the probe.
Meter fails to calibrate
NaCl.
Incorrect EC calibration.
Recalibrate the meter in EC
range. Set cell constant to 1.
At startup the meter displays
all LCD tags permanently
One of the keys is blocked.
Check the keyboard or con-
tact your dealer.
“Err xx” message displayed
at start up.
Internal error.
Contact your dealer or any
Milwaukee Instruments Service
Center.
3
www.milwaukeeinst.com
FUNCTIONAL DESCRIPTION
FRONT PANEL
1. Liquid Crystal Display (LCD)
2. CAL key, to enter/exit calibration
mode
3. GLP/ACCEPT key, to display GLP data
or to confirm value
4. SHIFT key, to select second key func-
tion
5. LOG/MR key, to store reading in
memory or to enter/exit recall mode
6. SETUP key, to enter/exit setup mode
or to toggle between delete one and
all logged data
7.
/ATC key, to manually decrease
temperature value or other parameters
or to select temperature compensation
mode
8. RANGE/FIXED key, to select measure-
ment range/switch focused data or to
freeze current reading on the LCD
9.
/TC key, to manually increase tem-
perature value or other parameters or
to view temperature coefficient value
10. ON/OFF key, to turn the meter ON
and OFF
11. Secondary LCD
12. Primary LCD
REAR PANEL
13. USB connector
14. RS232 connector
15. Power supply socket
16. DIN connector for probe