background image

Instruction Manual Mi 190 DO Bench Meter

7

PROBE CONNECTION  AND PREPARATION

To take measurements, connect the D.O. probe to the meter securely by aligning the

 

pins with the socket located on the back of the meter, pushing the plug in and tightening

 

the threaded ring.
Probes shipped from 

Milwaukee 

Instruments are dry. To hydrate the probe and 

prepare it

 

for use, connect it to the meter and proceed as follows:

1. Remove the red and black plastic cap. This cap is for

shipping purposes and can be thrown away.

2. Wet the sensor by soaking the bottom 2½ cm (1") of

the probe in electrolyte (MA 9071) for 5 minutes.

3. Rinse the membrane cap (MA 841     supplied in the kit

with the meter) with electrolyte solution while shaking it
gently. Refill with clean electrolyte solution.

4. Tap gently the sides of the membrane cap with your

finger tip to ensure that no air bubbles are trapped. To
avoid damaging the membrane, do not tap it directly on
the bottom.

5. Make sure that the rubber O-ring sits properly inside

the membrane cap.

6. With the sensor facing down, slowly screw the cap

clockwise. Some electrolyte will overflow.

When not in use and during polarization (see page 8), use
the protective transparent cap supplied in the kit with the
meter.

Shipping

cap

black

red

FILL FIRST

THEN TAP

THEN SCREW

BACK ON

2 2

2 2

2 2

2 2

2 2

TEMPERATURE CALIBRATION (for technical personnel only)

The Mi 190

Mi 190

Mi 190

Mi 190

Mi 190 is factory calibrated for temperature.

Martini’s DO probes are interchangeable and no temperature calibration is needed when they

 

are replaced.
If the temperature measurements are inaccurate, temperature recalibration should be performed.

 

For an accurate recalibration, contact your dealer or the nearest 

Milwaukee

 Customer Service 

Center,

 

or follow the instructions bellow.

• Prepare a vessel containing ice and water and another one containing hot water (at a

temperature of around 50 °C). Place insulation material around the vessels to minimize
temperature changes.

• Use a calibrated thermometer with a resolution of 0.1 °C as a reference thermometer.

• With the instrument off, press and hold down the LOG & MR keys, then power on the

instrument. The “CALIBRATION” tag will appear and the secondary LCD will show 0.0 °C.

• Immerse the temperature probe in the vessel with ice and water as near as possible to the

reference thermometer. Allow a few seconds for the probe to stabilize.

• Use the UP and DOWN arrow keys to set the reading on the secondary LCD to that of ice

and water, measured by the reference thermometer. When the reading is stable and close to
the selected calibration point, the “READY” and “ACCEPT” tags will blink.

• Press the ACCEPT key to accept the calibration point. The secondary LCD will show 50.0 °C.

• Immerse the temperature probe in the second vessel as near as possible to the reference

thermometer. Allow a few seconds for the probe to stabilize.

• Use the UP and DOWN arrow keys to set the reading on the secondary LCD to that of the

hot water, measured by the reference thermometer.

• When the reading is stable and close to the selected calibration point, the “READY” and

“ACCEPT” tags will blink.

• Press the ACCEPT key to accept the calibration point. The instrument returns to measurement

mode.

Note

Note

Note

Note

Note: If the reading is not close to the selected calibration point, “WRONG” tag will blink. Change

the temperature probe and restart calibration.

www.milwaukeeinst.com

Summary of Contents for Mi 190

Page 1: ...11 6726 Szeged Hungary Tel 36 62 428 050 Fax 36 62 428 051 e mail sales milwaukeeinst com Milwaukee Instruments Inc 2950 Business Park Drive Rocky Mount NC 27804 USA Tel 1 252 443 3630 Fax 1 252 443...

Page 2: ...NCTIONALDESCRIPTION B A C D E F G H I 27 WARRANTY WARRANTY WARRANTY WARRANTY WARRANTY This instrument is warranted against defects in materials and manufacturing for a period of 3 years from the date...

Page 3: ...12Vdc 13 14 2 6 2 6 2 6 2 6 2 6 ACCESSORIES MA 9070 Zero Oxygen Solution 230 ml MA 9071 Refilling Electrolyte Solution 30 ml MA 9311 115VAC to 12VDC converter MA 9310 230VAC to 12VDC converter MA 840...

Page 4: ......

Page 5: ...CAN is sent when the instrument is logging Err6 Err8 is sent when the command is incorrect or the instrument is not in measurement mode COMMANDS REQUIRING AN ANSWER The instrument will answer for the...

Page 6: ......

Page 7: ...e temperature recalibration should be performed For an accurate recalibration contact your dealer or the nearest Milwaukee Customer Service Center or follow the instructions bellow Prepare a vessel co...

Page 8: ...11 7 11 2 10 8 10 4 10 0 9 7 9 3 9 0 8 9 32 0 2 13 8 13 3 12 8 12 4 11 9 11 5 11 0 10 6 10 2 9 9 9 5 9 2 8 8 8 5 8 4 35 6 4 13 1 12 6 12 2 11 7 11 3 10 9 10 5 10 1 9 7 9 3 9 0 8 7 8 4 8 0 7 9 39 2 6...

Page 9: ...S e a L e v e l F 0 g l 1 0 g l 2 0 g l 3 0 g l 3 5 g l 0 1 4 6 0 1 3 6 4 1 2 7 4 1 1 9 0 1 1 5 0 3 2 0 2 1 3 8 1 1 2 9 1 1 2 0 7 1 1 2 9 1 0 9 1 3 6 5 4 1 3 0 9 1 2 2 5 1 1 4 7 1 0 7 3 1 0 3 8 3 9 2...

Page 10: ...e was stored and the temperature of the sample the longer the time will be Note If is displayed the D O probe is not properly connected or the temperature is out of range This also indicates the posib...

Page 11: ...witch the meter on by pressing the ON OFF switch For an accurate calibration it is recommended to wait for at least 15 minutes to ensure precise conditioning of the probe Remove the protective cap fro...

Page 12: ...ty factor The altitude factor The dEL message on the primary LCD and the record number on the secondary LCD along with ACCEPT tag blinking Notes Notes Notes Notes Notes If one of the arrow keys is pre...

Page 13: ...tores the value and adjusts the slope point The instrument stores the slope calibration data and returns to measurement mode Note Note Note Note Note If the reading is not close to the selected value...

Page 14: ...measurement mode e g 18 records free If there are less than 6 memory locations remaining the record number and Lo message will blink to alert the user 15 If the log space is full FULL LOC message wil...

Reviews: