Instruction Manual Mi 190 DO Bench Meter

3

FUNCTIONAL DESCRIPTION

FRONT PANEL

FRONT PANEL

FRONT PANEL

FRONT PANEL

FRONT PANEL

1. Liquid Crystal Display (LCD)

2. CAL key, to enter/exit calibration mode

3. MR key, to enter/exit recall mode

4. ACCEPT key, to confirm value

5. LOG key, to store reading in memory

6. SETUP key, to enter/exit setup mode or

to toggle between delete one and all
logged data

7.

key, to select SETUP item or to de-

crease parameter value

8. RANGE key, to select measurement unit/

switch focused data

9.

key, to select SETUP item or to

increase parameter value

10. ON/OFF key, to turn the meter ON

and OFF

11. Secondary LCD

12. Primary LCD

REAR PANEL

REAR PANEL

REAR PANEL

REAR PANEL

REAR PANEL

13. USB connector

14. RS232 connector

15. Power supply socket

16. DIN connector for probe

12

11

10

9
8
7
6

5

4

3

2

1

15

16

RS232

USB

Power
12Vdc

13

14

2 6

2 6

2 6

2 6

2 6

ACCESSORIES

MA 9070

Zero Oxygen Solution, 230 ml

MA 9071

Refilling Electrolyte Solution, 30 ml

MA 9311

115VAC to 12VDC converter

MA 9310

230VAC to 12VDC converter

MA 840/2

Spare probe with 2 meters (6.7') cable

MA 840/10

Spare probe with 10 meters (33') cable

MA 840/20

Spare probe with 20 meters  (67') cable

MA 841

5 spare membranes

MA 5200

Windows

®

 compatible software application

MA 9350

9-pin PC connection cable

TROUBLESHOOTING

SYMPTOMS

PROBLEM

SOLUTION

Reading fluctuates up and
down (noise).

DO probe not properly
connected.

Insert the probe.

Display shows DO reading
blinking.

Reading out of range.

Recalibrate the meter;
Check the sample is within
measurable range;

Meter fails to calibrate or
gives faulty readings.

Broken DO probe.

Replace the probe.

At startup the meter displays
all LCD tags permanently

One of the keys is blocked.

Check the keyboard or
contact your dealer.

"Err xx" message displayed at
start up.

Internal error.

Contact your dealer or any

 

Milwaukee

  Instruments 

Service

 

Center.

www.milwaukeeinst.com

Summary of Contents for Mi 190

Page 1: ...11 6726 Szeged Hungary Tel 36 62 428 050 Fax 36 62 428 051 e mail sales milwaukeeinst com Milwaukee Instruments Inc 2950 Business Park Drive Rocky Mount NC 27804 USA Tel 1 252 443 3630 Fax 1 252 443...

Page 2: ...NCTIONALDESCRIPTION B A C D E F G H I 27 WARRANTY WARRANTY WARRANTY WARRANTY WARRANTY This instrument is warranted against defects in materials and manufacturing for a period of 3 years from the date...

Page 3: ...12Vdc 13 14 2 6 2 6 2 6 2 6 2 6 ACCESSORIES MA 9070 Zero Oxygen Solution 230 ml MA 9071 Refilling Electrolyte Solution 30 ml MA 9311 115VAC to 12VDC converter MA 9310 230VAC to 12VDC converter MA 840...

Page 4: ......

Page 5: ...CAN is sent when the instrument is logging Err6 Err8 is sent when the command is incorrect or the instrument is not in measurement mode COMMANDS REQUIRING AN ANSWER The instrument will answer for the...

Page 6: ......

Page 7: ...e temperature recalibration should be performed For an accurate recalibration contact your dealer or the nearest Milwaukee Customer Service Center or follow the instructions bellow Prepare a vessel co...

Page 8: ...11 7 11 2 10 8 10 4 10 0 9 7 9 3 9 0 8 9 32 0 2 13 8 13 3 12 8 12 4 11 9 11 5 11 0 10 6 10 2 9 9 9 5 9 2 8 8 8 5 8 4 35 6 4 13 1 12 6 12 2 11 7 11 3 10 9 10 5 10 1 9 7 9 3 9 0 8 7 8 4 8 0 7 9 39 2 6...

Page 9: ...S e a L e v e l F 0 g l 1 0 g l 2 0 g l 3 0 g l 3 5 g l 0 1 4 6 0 1 3 6 4 1 2 7 4 1 1 9 0 1 1 5 0 3 2 0 2 1 3 8 1 1 2 9 1 1 2 0 7 1 1 2 9 1 0 9 1 3 6 5 4 1 3 0 9 1 2 2 5 1 1 4 7 1 0 7 3 1 0 3 8 3 9 2...

Page 10: ...e was stored and the temperature of the sample the longer the time will be Note If is displayed the D O probe is not properly connected or the temperature is out of range This also indicates the posib...

Page 11: ...witch the meter on by pressing the ON OFF switch For an accurate calibration it is recommended to wait for at least 15 minutes to ensure precise conditioning of the probe Remove the protective cap fro...

Page 12: ...ty factor The altitude factor The dEL message on the primary LCD and the record number on the secondary LCD along with ACCEPT tag blinking Notes Notes Notes Notes Notes If one of the arrow keys is pre...

Page 13: ...tores the value and adjusts the slope point The instrument stores the slope calibration data and returns to measurement mode Note Note Note Note Note If the reading is not close to the selected value...

Page 14: ...measurement mode e g 18 records free If there are less than 6 memory locations remaining the record number and Lo message will blink to alert the user 15 If the log space is full FULL LOC message wil...

Reviews: