8.1. LABOVAC 

Øe

ØA

ØEV

ØR

ØE

l

H

H1

L

 

 

LABOVAC 1 

Length 

(L) 

Width (l) 

Height (H) 

Without separator 

Height (H1) 

With separator 

without filtration

with filtration 

S26 R3 

1700 650  800  1050 

1450 

S51 R3 

1700 650  800  1050 

1450 

S126 R3 

1700 650  800  1250 

1550 

S161 R5 

1950 750 1100  1400 

1750 

S251 R5 

1950 980 1100  1500 

1850 

S325 R5 

1950 980 1100  1500 

2000 

8.2. 

LABOVAC 2 C2020 / MIL 

Øe

ØE

ØR

ØEV

ØA

H

H1

l

L

 

 

LABOVAC 2 

Length 

(L) 

Width (l) 

Height (H) 

without separator 

Height (H1) 

with separator 

without filtration

with filtration 

S26 R5 

1950 1000 1100  1200 

1450 

S51 R5 

1950 1000 1100  1200 

1450 

S126 R5 

1950 1000 1200  1400 

1750 

S161 R8 

2350 1050 1250  1500 

1900 

S251 R8 

2350 1050 1250  1600 

1950 

S325 R8 

2350 1050 1250  1600 

2100 

 

Summary of Contents for LABOVAC 1

Page 1: ...ACUUM UNITS SAFETY REGULATIONS Spare parts and general instructions of liquid ring vacuum pump This document must not be disclosed or reproduced without prior authorisation from MIL S Failure to respe...

Page 2: ......

Page 3: ...7 8 2 LABOVAC 2 C2020 MIL 7 8 3 POWER SUPPLY SEE ELECTRICAL CIRCUIT DIAGRAM ATTACHED 8 8 4 CONNECTIONS TO THE PNEUMATIC NETWORK 8 8 5 WATER SUPPLY OF VACUUM PUMPS 8 SECTION 9 COMMISSIONING 9 9 1 FIRST...

Page 4: ...NVIRONMENTAL PROTECTION symbol this pictogram reminds the user of the need to sort waste store it safely and dispose of it correctly during maintenance operations GENERAL REMARK If they are used as th...

Page 5: ...und with borosilicate and its filtration fineness is approximately 1 100th of a micron This cartridge passes the so called sodium flame test Penetration of bacteria and living micro organisms is well...

Page 6: ...y a mechanical vacuum switch These instruments measure pressures in RELATIVE terms and must be set according to where they are installed using the values in the table below N B for the CYCLIC 2020 or...

Page 7: ...ION 6 STAND BY MECHANICAL VACUUM SWITCH For the CYCLIC 2020 or MILLENIUM the analogue sensor is backed up by a mechanical vacuum switch fitted as a safety measure which must be set lower than the regu...

Page 8: ...tion C2VAL OPTIONAL regulation sensor connection C1VAL stand by vacuum switch Cyclic 2020 600 TEST A A B B C C CYCLIC 2020 300 Main switch Depth 250 For switch positions of the CYCLIC 2020 see the ele...

Page 9: ...50 S51 R3 1700 650 800 1050 1450 S126 R3 1700 650 800 1250 1550 S161 R5 1950 750 1100 1400 1750 S251 R5 1950 980 1100 1500 1850 S325 R5 1950 980 1100 1500 2000 8 2 LABOVAC 2 C2020 MIL e E R EV A H H1...

Page 10: ...1 5 1 5 1 5 2 5 2 5 4 Protection to be fitted A 4 6 10 16 16 25 For 2 pumps Cable section mm 1 5 2 5 2 5 4 6 10 Protection to be fitted A 6 10 16 25 32 50 8 4 Connections to the pneumatic network 8 4...

Page 11: ...bleed valve and open the network outlet valve Position all the pumps switches to On I and all the pumps will start up resulting in a steady increase of the depression within the network Check all con...

Page 12: ...SW2 4 and SW2 5 SECTION 11 ALARMS 11 1 For the CYCLIC 2020 When the LED on the key is lit an alarm has occurred constant Fault on concerned vacuum pump may be in connection with Circuit breaking disj...

Page 13: ...r mechanical pressure switch 12 1 Filtration IMPORTANT For this operation use all necessary protective clothing gloves mask chemical protective suit etc then destroy the contaminated consumable anti b...

Page 14: ...00 mbar Once the check has been completed set the general switch to O put the fuses back in place and then reset the general switch to I 12 4 Validation of sensor measurements vacuum gauge For CYCLIC...

Page 15: ......

Page 16: ...72 78 00 40 T l copie 33 0 4 78 00 82 34 N identification TVA FR 48327218939 SIRET 327 218 939 00079 RCS LYON B 327 218 939 CODE APE 2813 Z www mils fr e mail sales mils fr AGENCE ILE DE FRANCE 128 bi...

Reviews: