MIL'S LABOVAC 1 General Installation And Maintenance Instructions Download Page 13

General installations and maintenance instructions 

LABOVAC VACUUM UNITS 

 

Manual N° 514774 version 8 

05/2013 

Page 11/12

 

 

SECTION 12.  MAINTENANCE TABLE 

 
 

Monthly 

Quarterly 

Annually  2 years 

5 years 

FILTRATION 

-

 

bleed PYREX vessel 

-

 

change main anti-bacteria filter 

-

 

change secondary anti-bacteria filter (option) 

Regular inspection 

  # 

 

 

According to use 

MAIN and SECONDARY RECEIVERS 

-

 

general inspection of condition 

-

 

call in an outside inspection body 

  # 

 

 

According to legislation 

ELECTRONICS CABINET 

-

 

clean and dust using a soft brush

 

-

 

check the connections in the cabinet

 

-

 

check the connections at motor terminals 

 

-

 

operate the mechanical back-up system (LABOVAC2) 

-

 

check the intensity of the motors 

  # 

 

 

  # 

 

 

 

 

 

 

  # 

 

 

  # 

 

 

LIQUID-RING VACUUM PUMPS 

Refer to the start-up and maintenance manual 

VACUUM HOSE  

-

 

Check general condition 

 

-

 Change 

  # 

 

 

 

 

 

 

NON RETURN VALVE  

-

 

Check tightness  

 

 

 

 

PRESSURE INDICATOR 

(electronic sensor, vacuum gauge)

 

-

 

Check measures for correctness 

 

 

 

 

NETWORK ALARM  

-

 

Check correct working order  

 

 

 

 

EXHAUST HOSE  

-

 

Check the general condition 

 

-

 Change 

  # 

 

 

 

 

 

 

WATER SUPPLY 

-

 

Clean the screen of the filter 

 

 

 

 

-

 

Check solenoid valve 

 

 

 

 

-

 

Check safety of the water mechanical pressure switch 

 

 

 

 

12.1. Filtration 

IMPORTANT: 
For this operation, use all necessary protective clothing (gloves, mask, chemical protective 
suit, etc.) then destroy the contaminated consumable (anti-bacterial cartridges, liquid of 
condensation,…) according to applicable laws in your country (licensed incinerator). 

 
During and after maintenance, to respect basic hygiene rules. 
Before ordering tools used for maintenance, clean them with a disinfectant. 

12.1.1.  Bleed the PYREX vessel 

 

Position the three-way valve to bypass, 

 

Bring to atmospheric pressure by pressing the red button on the filter nose, 

 

Unscrew the vessel, empty it and screw it back into position. 

 

Put the three-way valve back into working position. 

12.1.2.  Changing the anti-bacteria filter 

Systematically replace the anti-bacteria cartridge every 12 MONTHS or more frequently if pressure on the network 
drops too low. 

 

Position the three-way valves to bypass, 

 

Bring to atmospheric pressure by pressing the red button on the filter nose, 

 

Unscrew the chamber, remove the white shell + filter element assembly; remove the collar holding the shell 
to the filter element. 

 

Fit the shell to the new filter element, replace the collar and clean the chamber. 

 

Then, once reassembled, position the three-way valves back to their working positions. 

Summary of Contents for LABOVAC 1

Page 1: ...ACUUM UNITS SAFETY REGULATIONS Spare parts and general instructions of liquid ring vacuum pump This document must not be disclosed or reproduced without prior authorisation from MIL S Failure to respe...

Page 2: ......

Page 3: ...7 8 2 LABOVAC 2 C2020 MIL 7 8 3 POWER SUPPLY SEE ELECTRICAL CIRCUIT DIAGRAM ATTACHED 8 8 4 CONNECTIONS TO THE PNEUMATIC NETWORK 8 8 5 WATER SUPPLY OF VACUUM PUMPS 8 SECTION 9 COMMISSIONING 9 9 1 FIRST...

Page 4: ...NVIRONMENTAL PROTECTION symbol this pictogram reminds the user of the need to sort waste store it safely and dispose of it correctly during maintenance operations GENERAL REMARK If they are used as th...

Page 5: ...und with borosilicate and its filtration fineness is approximately 1 100th of a micron This cartridge passes the so called sodium flame test Penetration of bacteria and living micro organisms is well...

Page 6: ...y a mechanical vacuum switch These instruments measure pressures in RELATIVE terms and must be set according to where they are installed using the values in the table below N B for the CYCLIC 2020 or...

Page 7: ...ION 6 STAND BY MECHANICAL VACUUM SWITCH For the CYCLIC 2020 or MILLENIUM the analogue sensor is backed up by a mechanical vacuum switch fitted as a safety measure which must be set lower than the regu...

Page 8: ...tion C2VAL OPTIONAL regulation sensor connection C1VAL stand by vacuum switch Cyclic 2020 600 TEST A A B B C C CYCLIC 2020 300 Main switch Depth 250 For switch positions of the CYCLIC 2020 see the ele...

Page 9: ...50 S51 R3 1700 650 800 1050 1450 S126 R3 1700 650 800 1250 1550 S161 R5 1950 750 1100 1400 1750 S251 R5 1950 980 1100 1500 1850 S325 R5 1950 980 1100 1500 2000 8 2 LABOVAC 2 C2020 MIL e E R EV A H H1...

Page 10: ...1 5 1 5 1 5 2 5 2 5 4 Protection to be fitted A 4 6 10 16 16 25 For 2 pumps Cable section mm 1 5 2 5 2 5 4 6 10 Protection to be fitted A 6 10 16 25 32 50 8 4 Connections to the pneumatic network 8 4...

Page 11: ...bleed valve and open the network outlet valve Position all the pumps switches to On I and all the pumps will start up resulting in a steady increase of the depression within the network Check all con...

Page 12: ...SW2 4 and SW2 5 SECTION 11 ALARMS 11 1 For the CYCLIC 2020 When the LED on the key is lit an alarm has occurred constant Fault on concerned vacuum pump may be in connection with Circuit breaking disj...

Page 13: ...r mechanical pressure switch 12 1 Filtration IMPORTANT For this operation use all necessary protective clothing gloves mask chemical protective suit etc then destroy the contaminated consumable anti b...

Page 14: ...00 mbar Once the check has been completed set the general switch to O put the fuses back in place and then reset the general switch to I 12 4 Validation of sensor measurements vacuum gauge For CYCLIC...

Page 15: ......

Page 16: ...72 78 00 40 T l copie 33 0 4 78 00 82 34 N identification TVA FR 48327218939 SIRET 327 218 939 00079 RCS LYON B 327 218 939 CODE APE 2813 Z www mils fr e mail sales mils fr AGENCE ILE DE FRANCE 128 bi...

Reviews: