background image

 

 

 

Sterylis VS Manual                                                                                                                                                                                       

| 25

 

 

To activate the front panel locking mode: 

During active standby, silent disinfection, standard disinfection or filtration, press and hold the + and - buttons in the O

STERILIZATION section for at least 3 seconds.  
 

To deactivate the front panel locking mode: 

While the front panel lock mode is active, press and hold the + and - buttons in the O

STERILIZATION section for at least 3 

seconds.  

6.6 Changing the configuration parameters 

To change the configuration parameters, press and hold the  SET button in the  OZONING TIME section in the standby 
mode. Then enter the appropriate code to change the parameters. 
 

Entering the access code: 

The 4-digit code entered is shown on the display under O

3

 STERILIZATION. The + and - buttons change the value by 1 

active  digit.  An  active  number  is  the  one  that  flashes  at  a  high  frequency,  inactive  numbers  light  up  constantly.  The 
PROG.SET  button  changes  the active  digit  to  the  next  one  on  the  right.  After  entering  the  correct  code,  confirm  it  by 
pressing the yellow START button longer. If the code is incorrectly entered, the code XXXX or HHHH will be displayed and a 
short beep will occur. In this case, the code must be re-entered. 
 
You can change the following parameters: 

Parameter 
code 

Description 

Min value 

Max value 

Default value 

C101 

Operating concentration in AUTO mode [ppm] 

10 

C130 

Siren operation in sterilisation mode [ON/OFF] 

C138 

50% reduction in the power of ozone generators 

C150 

Date setting - day of the month 

31 

15 

C151 

Date setting - month 

12 

C152 

Date setting - year 

2020 

2120 

2020 

C200 

FIRMWARE revision (read only) 

FW_REV 

The  individual  parameters  are  changed  using  the  +  and  -  buttons  in  the  O

STERILIZATION  section.  The  code  of  the 

selected parameter is displayed in the O

STERILIZATION section. The current value of the selected parameter is displayed 

in  the  O

CONCENTRATION  section.

 

To  change  the  value  of  the  parameter,  operate  the  +  and  -  buttons  in  the  O

CONCENTRATION. The active digit is selected using the SET button in the same section of the user panel. To confirm the 
value of the selected configuration parameter, press and hold the PROG.SET button. The selected value will be confirmed 
by a short beep and all active displays will flash. 
 
To return to the standby mode, press and hold the SET button in the OZONING TIME section. 

6.7 Working time counters 

 

The device is equipped with the operating time counters listed in the table below. The user has the possibility to read the 
indications of these counters and reset them.  
 
To view or reset the counters, press and hold the SET button in the OZONING TIME section while the unit is in standby 
mode. Then enter the appropriate code to change the parameters. 
 

Entering the access code: 

The 4-digit code entered is shown on the display under O

3

 STERILIZATION. The + and - buttons change the value by 1 

active  digit.  An  active  number  is  the  one  that  flashes  at  a  high  frequency,  inactive  numbers  light  up  constantly.  The 
PROG.SET  button  changes  the active  digit  to  the  next  one  on  the  right.  After  entering  the  correct  code,  confirm  it  by 

Summary of Contents for STERYLIS VS-100/450

Page 1: ...MODELS VS 100 450 VS 160 600 VS 300 900 VS 400 1200 VS 600 1500 VS 900 1800 MANUAL USER S PROFESSIONAL ROOM STERILISERS...

Page 2: ...with the ozone concentration limits 23 6 3 4 5 Emergency situations during the sterilization process 24 6 4 Sleep mode 24 6 5 Front panel locking mode 24 6 6 Changing the configuration parameters 25...

Page 3: ...inside the device is high and can cause an electric shock This is especially true for the very high voltage generated by the ozone generators It is dangerous to touch any parts inside the device In th...

Page 4: ...arning icons They are located in the manual on the housing of the device or on other parts of the steriliser and are directly related to it Symbol Meaning Symbol Meaning Important safety instruction B...

Page 5: ...pale cold skin sweat chills accelerated heart rate place them in an anti shock position on their back with their legs raised Despite a specially designed disinfection channel and limiting switch prot...

Page 6: ...pain tearing itching redness avoid direct sunlight cool your eyes with a cloth dampened with cold water or glasses with cooling gel Do not scratch or rub your eyes If you have contact lenses it is ess...

Page 7: ...s disease and increased ailments in people with coexisting respiratory and cardiovascular diseases In very high concentrations it is dangerous to live The effect of ozone on living organisms depending...

Page 8: ...es avoid contact with eyes Wash your hands thoroughly with water and soap after providing medical assistance Inhalation If the airways come into contact with too high a concentration of ozone leave th...

Page 9: ...eir indications are correct Improper measurements of the ozone sensor can lead to life threatening and health threatening effects as described in this chapter The ozone sensor is calibrated by the man...

Page 10: ...dance with EN 136 and EN 14387 standards and in the skin protection equipment in accordance with the personal protective equipment table When the ozoning of the room is complete it must be thoroughly...

Page 11: ...in the room including electronics it is also advisable to move them during disinfection The use of ozone sterilisation ozoning from time to time should not affect the destruction of objects in the roo...

Page 12: ...ria and allergens using a coating applied to the filter medium In addition to particles the active carbon filter retains harmful gases and odours through adsorption In the UV C disinfection mode the U...

Page 13: ...on dose 149 J m2 232 J m2 443 J m2 Ozone generator capacity 30 000 mg h 45 000 mg h 60 000 mg h Net weight 18 kg 20 kg 35 kg in sterilisation mode Name STERYLIS STERYLIS STERYLIS Model VS 400 1200 VS...

Page 14: ...ts with other molecules breaking down their structures The effectiveness of ozone in eliminating odours lies in the fact that as a gas it is able to reach wherever there is air i e all the cracks nook...

Page 15: ...pecially selected type and high power of the UV C light source with a selective emitted wavelength of 253 7 nm and the unique solution of the double quartz filter UV C lamp used in STERYLIS steriliser...

Page 16: ...g manual 6 Safety data sheet ozone 7 UV C safety data sheet 8 Safety data sheet 9 Declarations of conformity in Polish and English 10 Two warning signs WARNING NO ENTRY STERILISATION IN PROGRESS as be...

Page 17: ...front of the device The user panel consists of a main switch controls individual modes of operation failure and warning function keys three segment displays and a graphical display of the ozone diagr...

Page 18: ...ONCENTRATION section segment display indicating the ozone concentration and the and SET buttons Alarm icons for failure of a system of ozone generators ozone sensor or UV C fluorescent lamps Main devi...

Page 19: ...enter the standby mode and then display the corresponding error codes on the display in the O3 STERILIZATION section except for errors F201 and F202 which cause the PRE TEST to stop until the error is...

Page 20: ...PRE TEST function and then it will start to light steadily and the device will start working in the selected mode of operation Confirmation of the disinfection mode is the operation of the front panel...

Page 21: ...where the user can change the parameters of ozoning time and working ozone concentration 6 3 4 1 O3 sterilisation automatic mode To start O3 STERILIZATION in automatic mode press and hold the yellow S...

Page 22: ...igh frequency inactive numbers light up constantly The PROG SET button changes the active digit to the next one on the right After entering the correct code confirm it by pressing the yellow START but...

Page 23: ...During the ozone destruction process the LED backlight panel is activated When the safe concentration 0 1 ppm is reached the ozone safety light will light up After completion of the sterilization pro...

Page 24: ...ower supply of the device should be disconnected as safely as possible preferably by means of an overcurrent switch in the switchgear or if necessary by unplugging the plug of the power cord from the...

Page 25: ...generators 0 1 0 C150 Date setting day of the month 1 31 15 C151 Date setting month 1 12 7 C152 Date setting year 2020 2120 2020 C200 FIRMWARE revision read only FW_REV The individual parameters are c...

Page 26: ...value of the electricity consumed is previewed using the d201 meter 6 8 Other functionality of the device Other remaining functionality of the device is described in subsections 6 8 1 6 8 2 and 6 8 3...

Page 27: ...e sensor error has been detected during previous use of the unit the OZONE SENSOR TEST function can erase this error if the unit will function correctly during this function applies to the following e...

Page 28: ...nd setting the current date The function is used to set the current time on the controller Selecting the SET TIME function displays the currently set time on the display in the OZONING TIME section To...

Page 29: ...1 and 7 2 2 Do not let the filters clog and avoid their excessive operation After the service life of the UV lamps 9 000 operating hours in the steriliser they should be replaced This is related not...

Page 30: ...t air filter cartridge towards the floor 4 Remove the air filter from the cartridge 5 Insert a new air filter into the cartridge making sure that the direction of the arrow on the cartridge and on the...

Page 31: ...of the end point also results in a corresponding error on the display in the O3 STERILIZATION section If the end point violation stops the error is automatically reset The safety limiting switches pr...

Page 32: ...s time has been exceeded will be displayed Note The replacement of UV C fluorescent tubes in STERYLIS sterilisers may only be carried out by the manufacturer s service or another authorised service ce...

Page 33: ...STERILIZATION mode according to the instructions in section 6 3 4 9 Exit the room by closing the door behind you and make sure again that it is tightly sealed and that no one is in the room immediatel...

Page 34: ...vice life of air filters F003 Exceeded service life of ozone generators F004 Exceeded service life of the ozone sensor F005 Exceeded expiry time of the ozone sensor F006 Air filters expiration date ex...

Page 35: ...t blinking PRE TEST function is in progress for standard disinfection mode active errors Yellow pictograms O3 and flashing PRE TEST function for sterilization mode in progress no active errors Yellow...

Page 36: ...use within certain voltage limits Directive 2014 30 EU on the harmonisation of the laws of the Member States relating to electromagnetic compatibility Directive 2011 65 EU on the restriction of the u...

Page 37: ...t with the service department Phone 48801080 257 E mail address serwis emiloo pl When contacting the service department you may be required to provide the data on the unit s nameplate Serial number Da...

Reviews: