ILE-61 Sensing Head – INSTRUCTION MANUAL
7ML19981CX01
Integrator Calibration
Refer to the flowmeter integrator instruction manual for integrator calibration
instructions.
To apply the Test Weight for the Span calibration:
1.
Mount the calibration bracket (stored on the left side of the cast frame).
2.
Attach one end of a string (monofilament fishing line or fine flexible cable) to the
Test Weight.
3.
Route the other end of string over the calibration pulley.
4.
Attach the free end of the string to the string fixing bolt. Ensure the string rests in the
bolt groove.
5.
When the calibration is complete, return the calibration bracket to the storage
position.
Note:
Ensure the Test Weight is suspended free from obstruction.
string fixing bolt
calibration arm
test weight
Note:
Accurate calibration is not assured until material tests and a Span Adjust are
performed.
376draft1.fm Page 12 Wednesday, November 21, 2001 1:54 PM