![Millipore MERCK EZ-Pak User Manual Download Page 26](http://html1.mh-extra.com/html/millipore/merck-ez-pak/merck-ez-pak_user-manual_1796895026.webp)
Cleaning the EZ-Pak® Dispenser Curve
1. Switch off and disconnect the EZ-Pak® Dispenser Curve from
the power outlet before cleaning.
2. Wipe outside surfaces of the dispenser with a non-woven wipe
moistened with one of the products listed in the following table.
NOTE
Do not spray any cleaning products, solvents, or abrasives onto or
into the EZ-Pak® Dispenser Curve, as these could damage the
mechanism.
Cleaning Product
Validated
dilution
1
Ethanol (<30%)
Ready to use
Didecyldimethylammonium chloride (<1%)
Propanol (<1%)
N-(3-aminopropyl)-N-dodecylpropane-1,3-diamine (<1,5%)
2
Isopropyl alcohol 70%
Ready to use
3
Ethanol 70%
Ready to use
4
Sodium hypochlorite
250 ppm
5
Peracetic acid at (5 to 15%)
2.5%
Acetic acid at (5 to 15%)
Hydrogen peroxide (15 to 30%)
6
Ethanol (25 to 50%)
Ready to use
Polyhexamethylene biguanide hydrochloride (<2,5%)
N-(3-aminopropyl)-N-dodecylpropane-1,3-diamine (<2,5%)
Didecyldimethylammonium chloride (<1%)