Millipore Amicon Ultra-4 Series User Manual Download Page 6

Amicon

®

 Ultra-4 Centrifugal Filter Devices

6

Chemical Compatibility

Amicon

®

 Ultra centrifugal devices are intended for use with biological fluids and aqueous solutions. 

 

Before use, check the sample for chemical compatibility with the device.

Table 6. Chemical Compatibility of Amicon

®

 Ultra Filter Devices

Acids

Concentration

Concentration

Acetic acid 

≤ 50%*

Phosphoric acid 

≤ 30%

Formic acid 

≤ 5%*

Sulfamic acid

≤ 3%

Hydrochloric acid 

≤ 1.0 M

Sulfuric acid 

≤ 3%

Lactic acid 

≤ 50%

Trichloroacetic acid (TCA)

≤ 10%*

Nitric acid 

≤ 10%

Trifluoroacetic acid (TFA)

≤ 30%*

Alkalis

Ammonium hydroxide 

≤ 10%

Sodium hydroxide 

≤ 0.5 M

Alcohols

n-Butanol 

≤ 70%

Isopropanol 

≤ 70%

Ethanol 

≤ 70%

Methanol 

≤ 60%

Detergents

Alconox

®

 detergent 

≤ 1%

Sodium dodecyl sulfate (SDS)

≤ 0.1%

CHAPS detergent

≤ 0.1%

Tergazyme™

 

detergent 

≤ 1%

Lubrol

®

 PX detergent

≤ 0.1%

Triton

®

 X-100 surfactant

≤ 0.1%

Nonidet

®

 P-40 surfactant

≤ 2%

Tween

®

 20 surfactant

≤ 0.1%

Sodium deoxycholate

≤ 5%

Organic solvents

Acetone

not recommended

Ethyl acetate

not recommended

Acetonitrile 

≤ 20%

Formaldehyde 

≤ 5%

Benzene

not recommended

Pyridine

not recommended

Carbon tetrachloride

not recommended

Tetrahydrofuran

not recommended

Chloroform

not recommended

Toluene

not recommended

Dimethyl sulfoxide (DMSO)

≤ 5%*

Miscellaneous

Ammonium sulfate 

Saturated 

Phenol

≤ 1%

Diethyl pyrocarbonate 

≤ 0.2%

Phosphate buffer (pH 8.2)

≤ 1 M

Dithiothreitol (DTT)

≤ 0.1 M

Polyethylene glycol

≤ 10%

Glycerine

≤ 70%

Sodium carbonate 

≤ 20%

Guanidine HCl 

≤ 6 M

Tris buffer (pH 8.2)

≤ 1 M

Imidazole

≤ 100 mM

Urea 

≤ 8 M

Mercaptoethanol

≤ 0.1 M

* Contact with this chemical may cause materials to leach out of the component parts. Solvent blanks are recommended to determine 

whether leachables represent potential assay interferences.

Summary of Contents for Amicon Ultra-4 Series

Page 1: ...ncludes 5 different cutoffs MWCO Amicon Ultra 3K device 3 000 MWCO Amicon Ultra 10K device 10 000 MWCO Amicon Ultra 30K device 30 000 MWCO Amicon Ultra 50K device 50 000 MWCO Amicon Ultra 100K device...

Page 2: ...r typical spin times 4 To recover the concentrated solute insert a pipettor into the bottom of the filter device and withdraw the sample using a side to side sweeping motion to ensure total recovery T...

Page 3: ...ntrate Volume vs Spin Time Swinging bucket rotor Concentrate volume L Spin time min 3K device 10K device 30K device 50K device 100K device 10 1 369 176 73 32 264 15 76 46 36 20 478 58 37 30 33 25 228...

Page 4: ...ging bucket Fixed angle Cytochrome c 0 25 mg mL 3K 40 94 62 43 5 65 0 98 2 96 7 Cytochrome c 0 25 mg mL 10K 15 76 54 52 3 76 6 97 3 98 5 BSA 1 mg mL 30K 10 73 42 56 1 98 6 95 8 95 0 BSA 1 mg mL 50K 10...

Page 5: ...terial filtrate and concentrate to determine solute concentration 8 Calculate recoveries using the weight volume data and the measured concentrations as follows concentrate recovery 100 Wc Cc Wo Co fi...

Page 6: ...ent 0 1 Tergazyme detergent 1 Lubrol PX detergent 0 1 Triton X 100 surfactant 0 1 Nonidet P 40 surfactant 2 Tween 20 surfactant 0 1 Sodium deoxycholate 5 Organic solvents Acetone not recommended Ethyl...

Page 7: ...o diagnostic IVD use All other devices are for research use only Notice We provide information and advice to our customers on application technologies and regulatory matters to the best of our knowled...

Reviews: