Millipore Amicon Ultra-4 Series User Manual Download Page 2

Amicon

®

 Ultra-4 Centrifugal Filter Devices

2

Suitability

Preliminary recovery and retention studies are suggested to ensure suitability for intended use. See the “How to 

 

Quantify Recoveries” section.

Rinsing Before Use

The ultrafiltration membranes in Amicon

®

 Ultra-4 devices contain trace amounts of glycerine. If this material 

 

interferes with analysis, rinse the device with buffer or Milli-Q

®

 water before use. If interference continues, rinse with 

0.1 N NaOH followed by a second spin of buffer or Milli-Q

®

 water.

CAUTION:  

Do not allow the membrane in Amicon

®

 Ultra filter devices to dry out once wet. If you are not using the 

device immediately after rinsing, leave fluid on the membrane until the device is used.

How to Use Amicon

®

 Ultra-4 Centrifugal Filter Devices

1.  Add up to 4 mL of sample (3.5 mL if using a 23° fixed-angle rotor) to the Amicon

®

 Ultra filter device.

2.  Place capped filter device into centrifuge rotor (swinging-bucket preferred); counterbalance with a similar device.
3. 

 

When using a swinging-bucket rotor

, spin the device at 4,000 × g maximum for approximately 10–40 minutes.

When using a fixed-angle rotor, orient the device with the membrane panel facing up.
For Amicon

®

 Ultra 3K, 10K, 30K, and 50K devices

, spin at 7,500 × g maximum for approximately 10–40 

 

minutes.

For Amicon

®

 Ultra 100K devices

,

 spin at 5,000 × g maximum for approximately 10–20 minutes.

NOTE: 

Refer to Figures 1 and 2, and Tables 2 and 3 for typical spin times.

4.   To recover the concentrated solute, insert a pipettor into the bottom of the filter device and withdraw the sample 

using a side-to-side sweeping motion to ensure total recovery. The ultrafiltrate can be stored in the centrifuge tube.

NOTE:

 

For optimal recovery, remove concentrated sample immediately after centrifugation.

Desalting or Diafiltration

Desalting, buffer exchange, or diafiltration are important methods for removing salts or solvents in solutions 

containing biomolecules. The removal of salts or the exchange of buffers can be accomplished in the Amicon

®

 Ultra-4 

device by concentrating the sample, then reconstituting the concentrate to the original sample volume with any 

desired solvent. The process of “washing out” can be repeated until the concentration of the contaminating microsolute 

has been sufficiently reduced. See example below.

Spin to 

 

concentrate

4 mL of

 

1 mg/mL

 

protein in

 

100 mM

 NaCl

50 µL of 

 

80 mg/mL

 

protein in 

 

100 mM NaCl

100 mM 

NaCl

50 µL of 

 

80 mg/mL 

protein in 

 

11.1 mM NaCl

11.1 mM 

NaCl

Spin to 

 

concentrate

Add 3.95 mL of 

10 mM NaCl or 

exchange buffer

4 mL of 

 

1 mg/mL 

 

protein in 

 

11.1 mM NaCl

Performance - DNA Concentration

We have determined that the Amicon

®

 Ultra-4 30K device provides the best balance between recovery and spin time 

for double-stranded DNA for base pairs ranging from 137 to 1,159.

Table 1. Typical Recovery of Nucleotides from Amicon

®

 Ultra-4 30K Device

Double-stranded

 

DNA Base Pair 

Size

Spin Time 

(min)

Concentrate  

Volume (µL)

Recovery (%)

137–1,159

10

50–70

> 85

Spin conditions: Fixed-angle rotor, 5,000 × g, room temperature, 2 mL starting volume.

Summary of Contents for Amicon Ultra-4 Series

Page 1: ...ncludes 5 different cutoffs MWCO Amicon Ultra 3K device 3 000 MWCO Amicon Ultra 10K device 10 000 MWCO Amicon Ultra 30K device 30 000 MWCO Amicon Ultra 50K device 50 000 MWCO Amicon Ultra 100K device...

Page 2: ...r typical spin times 4 To recover the concentrated solute insert a pipettor into the bottom of the filter device and withdraw the sample using a side to side sweeping motion to ensure total recovery T...

Page 3: ...ntrate Volume vs Spin Time Swinging bucket rotor Concentrate volume L Spin time min 3K device 10K device 30K device 50K device 100K device 10 1 369 176 73 32 264 15 76 46 36 20 478 58 37 30 33 25 228...

Page 4: ...ging bucket Fixed angle Cytochrome c 0 25 mg mL 3K 40 94 62 43 5 65 0 98 2 96 7 Cytochrome c 0 25 mg mL 10K 15 76 54 52 3 76 6 97 3 98 5 BSA 1 mg mL 30K 10 73 42 56 1 98 6 95 8 95 0 BSA 1 mg mL 50K 10...

Page 5: ...terial filtrate and concentrate to determine solute concentration 8 Calculate recoveries using the weight volume data and the measured concentrations as follows concentrate recovery 100 Wc Cc Wo Co fi...

Page 6: ...ent 0 1 Tergazyme detergent 1 Lubrol PX detergent 0 1 Triton X 100 surfactant 0 1 Nonidet P 40 surfactant 2 Tween 20 surfactant 0 1 Sodium deoxycholate 5 Organic solvents Acetone not recommended Ethyl...

Page 7: ...o diagnostic IVD use All other devices are for research use only Notice We provide information and advice to our customers on application technologies and regulatory matters to the best of our knowled...

Reviews: