Maintenance and settings
16
Purging air from the IPAK Quanta ICP
®
cartridge
1: Make sure you are not in maintenance mode.
Otherwise, press the “Exit maintenance” button
to go back to the home screen.
2: Get a screwdriver with a very thin shaft and
small head. You can alternatively use a mechani
-
cal pencil provided that the lead is retracted.
3: Locate the small hole on the top part of the
IPAK Quanta ICP
®
cartridge.
3: Gently insert the screwdriver into the IPAK
Quanta ICP
®
cartridge. This will open a small
vent that will expel any air in the IPAK Quanta
ICP
®
cartridge.
Important: Be very careful when pushing the tip of the screwdriver into the vent hole of the IPAK
Quanta ICP
®
cartridge. The vent hole only needs to be opened a very small amount. Wear goggles.