Maintenance and settings
10
Consumables kit replacement standard operating procedure
1: Unpack the IPAK Quanta ICP
®
Cartridge. Write today’s date on the small sticker at the bottom.
2: Remove the Milli-Q
®
IQ Element unit front cov
-
er.
3: Click on the Menu button to access the
MAINTENANCE section.
4: Click on “Install consumables”.
5: Depending upon the system type that feeds
the Milli-Q
®
IQ Element unit, either click on “IPAK
Meta - IPAK Quanta - IPAK Quanta ICP -
Optimizer” or “IPAK Quanta - IPAK Quanta ICP -
Optimizer - installation”.
6: Depending upon the system type that feeds the Milli-Q
®
IQ Element unit, you now see either of
both screens shown below. Be sure to meet the pre-requisite ahead of pressing “Next”.
Important: In addition to the 20 L (Milli-Q
®
IQ 7003/7005/7010/7015) or 30 L (Milli-Q
®
IQ 7000) in the tank
required to proceed the wizard, additional 20L will be necessary for the Optimizer™ LW final filter rinsing (step
21). If your Milli-Q
®
IQ Element unit is fed by a Milli-Q
®
IQ 7003/7005/7010/7015 and a 25L capacity tank, you
will need to wait several hours for the tank to refill in order to finish the procedure depending on system flowrate
capacity.