Requirements and specifications
27
Feed Water
The system has been designed to operate within feed water requirements below:
Parameter
Value or range
Pressure
< 6 bar
(for pressures below 0 bar, the system will operate, but
product flow rate may be lower)
Feed water quality
Pretreated water, using EDI, DI, RO or distillation
technologies
Temperature
5 - 35°C (41 - 95°F)
Conductivity
< 50 μS/cm @25° C (FCE)
Total Organic Carbon (TOC)
< 50 ppb
System specifications
Electrical
The power supply converts mains voltage to 28 V. The power supply is compatible worldwide
.
Catalog
Frequency
Max power
Voltage
ZEQ7000T0
&
ZEQ7000TR
50-60 Hz ±2Hz
170 VA
100 – 240 Vac ± 10%
Environmental
Specific environmental conditions have to be respected to ensure normal operation of the
Milli-Q system.
Location
indoor use only
Ambient operating temperature
room temperature ranges from 4°C to 40°C
Relative humidity
maximum relative humidity of 80% for a
temperature up to 31°C, decreasing linearly to a
relative humidity of 50% at 40°C
Altitude
up to 3000m above sea level
Installation category
bench / underbench or wall mounted
Pollution degree
2
Noise level
< 50 dB at 1m