Settings

13

System configuration

Lab close

When leaving the lab for long periods of inactivity, this mode can be activated to save energy 

and reduce wear of system components. Recirculation is reduced to once a day. When it is 24 

hours prior to resuming lab activity, the system automatically resumes a recirculation every 

hour, ensuring it is ready for use.
1. Activate Lab close via the slider button.

2. Enter a Start and End date (minimum of 3 days). Lab close mode will be initiated at 00:01 

on the selected date.

 

Safety flow stop

 

 

A precautionary measure that stops a Q-POD dispensing after having continuously delivered 

water for a certain duration.

 

Adjust to the preferred time by using the arrows or click on the box to access the keypad.

By default the timer is set to 15 minutes and is active. This setting will not impact volumetric 

dispensing functions.

Screensaver 

 

This is the maximum duration of inactivity on the HMI before the screensaver starts.

 

1. Activate / deactivate via the slider button.

2. Adjust to the preferred time by using the arrows or click on the box to access the keypad.

By default the screensaver is active and the timer is set to 5 minutes.

Custom Volume 

 

This is the volume you can customize to be able to dispense with the Volumetric Dispensing 

function.

 

The volume can be set from 100 to 25000 mL in 100 mL increments.

HMI configuration

This app allows a user to configure all parameters specific to the HMI.

 

Screen brightness 

Adjust to the preferred brightness from 1 to 7 using the arrows or click on the box to access 

the keypad and type-in the value.

 

Buzzer 

The POD can emit a sound when alerts/alarms are triggered. This can be activated or 

deactivated using the slider button. By default, the sound is deactivated.

Summary of Contents for EQ 7000

Page 1: ...User Manual Milli Q EQ 7000...

Page 2: ...vity 15 Date Time Language 16 Units 17 INFORMATION MENU 18 System 18 History 18 Flow schematic 19 Troubleshooting 19 Contacts 19 MAINTENANCE MENU 20 Depressurization 20 Consumable installation wizards...

Page 3: ...e The Milli Q EQ 7000 is intended to produce ultrapure water from a purified water source primarily for use in a variety of laboratories worldwide The product is designed to produce ultrapure water wi...

Page 4: ...ed of ion exchange resins removing the bulk of the ions The purified water then flows to the UV 185nm oxidation lamp Organic molecules are partially oxidized resulting in ions These can then be captur...

Page 5: ...ter within the system will recirculate for three minutes every hour This is to maintain a clean environment and to avoid contamination The system should not be turned off as this will stop the periodi...

Page 6: ...ing 2 The system is powered on STEADY The water quality fits the requirements water quality values are below alarm setpoints BREATHING The system is in recirculation mode A water quality alarm is rais...

Page 7: ...ispensing To do this press on the recirculation icon this also refreshes the water quality parameters Product resistivity TOC Temperature There are several options to dispense ultrapure water Manual D...

Page 8: ...ed Water Quality values When the dispensing is finished the resistivity and temperature are instantly displayed on the left side of the HMI Ninety seconds later a new TOC value at the point of distrib...

Page 9: ...pecific alert or alarm to get further information and better understand the root cause Advice is provided in the Information tab to help the user troubleshoot If the issue is not resolved select the C...

Page 10: ...tions can easily be reached in a few clicks or swipes similar to a smartphone HOME SCREEN MAIN MENUS Home Water quality information Dispensing function Alerts Alarms when applicable Menu Consumables I...

Page 11: ...l automatically appear To remove the keypad click on or on any zone outside the keypad Note When in a menu or app level screen 1 minute of inactivity generates a timeout back to the home screen When i...

Page 12: ...information and a description of the purification technology Note Certificates of quality are now available online www mymilliqconsumables com All cartridges have been designed to be easily replaced b...

Page 13: ...d only Feed quality measure Read only Alarm relay activation Read only Alarm relay configuration Custom Volume HMI configuration Brightness Buzzer Read only Alarm configuration Product resistivity Rea...

Page 14: ...Note By default this parameter is activated The default password is PASS If the system manager password is forgotten contact the technical support hotline Log off as a Manager 1 Press on the menu butt...

Page 15: ...15 minutes and is active This setting will not impact volumetric dispensing functions Screensaver This is the maximum duration of inactivity on the HMI before the screensaver starts 1 Activate deactiv...

Page 16: ...f bad water quality is detected an alarm that stops water production will appear informing the user that the system is at risk and to check the feed water quality This will protect the downstream puri...

Page 17: ...s should be changed when the system alerts the user Saturated pack will start to let ion and organic contaminants through This is called breakthrough and often starts with traces that cannot be detect...

Page 18: ...ns are shown in the form 2020 07 06 YYYY MM DD YYYY indicates a four digit year MM indicates a two digit month of the year 01 through 12 DD indicates a two digit day of that month 01 through 31 Start...

Page 19: ...rature compensated resistivity or conductivity due to the fact that these two parameters are not measured exactly at the same time TC2 The actual temperature compensated resistivity or conductivity va...

Page 20: ...ng events report Each row is a dispense operation It is an archive of all the dispense reports To export data or preview a graph 1 Select a time period of interest with a start and end date 2 A graph...

Page 21: ...ffective troubleshooting when in contact with the technical support hotline System information last service visit Water quality consumables installed dates Active alerts alarms autotest results Diagno...

Page 22: ...able in the applicable wizards Maintenance screens Maintenance ongoing from the HMI Depressurization This temporarily stops production and all water processes This function is not necessary during nor...

Page 23: ...artridges may not be fully rinsed The remaining rinse will need to be performed manually Important During the physical pack replacement in the water purification unit both cartridges must first be rem...

Page 24: ...ion will ensure accuracy of your dispensing Note You will need a 1 0L graduated cylinder Select Care Cleaning app and click on Volumetric dispensing calibration Follow the self guided wizards UV lamp...

Page 25: ...rom its power source Installation and maintenance should only be done by a qualified person Appropriate personal protective equipment PPE must be worn and safe work practices must be followed If the s...

Page 26: ...ther cartridges installation IPAK other cartridges installation new cartridges not recognized Important During the physical replacement in the system unit both cartridges must first be removed before...

Page 27: ...Dispensing Volumetric dispensing System manager logged in Start dispense Stop dispense 1 Alarm with number of active alarm s Alert Home Menu Storage tank Back Mass storage USB connected Ethernet LAN s...

Page 28: ...installed and used in a laminar flow hood Pyrogens endotoxins 0 001 EU mL with Biopak filter RNases 2 pg mL with Biopak filter DNases 5 pg mL with Biopak filter Proteases 0 15 g mL with Biopak filter...

Page 29: ...er supply converts mains voltage to 28 V The power supply is compatible worldwide Catalog Frequency Max power Voltage ZEQ7000T0 ZEQ7000TR 50 60 Hz 2Hz 170 VA 100 240 Vac 10 Environmental Specific envi...

Page 30: ...d USB keys only work when formatted FAT32 NTFS format is not compatible Ethernet When connected through an Ethernet protocol the display interface can be accessed remotely using internet web browsers...

Page 31: ...ssories Note For wall mounting installation use a concrete wall capable of support max load of 16 Kg Use 6mm diameter stud and fasteners capable to support max load of 4 Kg each Recycling Directive 20...

Page 32: ...l ZEQ7C0NDC Internal Pressure Regulator ZF3000810 e Sure dispenser tip ZFDISTPA1 Alarm relay cable ZMQ0ALCA1 Consumables order at www mymilliqconsumables com Description Catalog Number IPAK Meta IPAK...

Page 33: ...ore Milli Q Q POD IPAK Meta IPAK Quanta Jetpore IQnano Millipak Biopak EDS Pak VOC Pak and LC Pak are trademarks of Merck KGaA Darmstadt Germany Millipore SAS is an affiliate of Merck KGaA Darmstadt G...

Page 34: ...r could exist This CAUTION sticker is used to refer to a surface that can be hot Disconnect and switch power off to allow surface to cool before servicing Hazard The Milli Q system must be connected t...

Reviews: