![MIHM-VOGT 7203000100 Operating Manual Download Page 63](http://html1.mh-extra.com/html/mihm-vogt/7203000100/7203000100_operating-manual_1793739063.webp)
Installation
Version 2.5en - 12/2020
MV-R 100
Page 17
Installation
Positioning
The
MV-R 100
is designed as a table-top device. For stable positioning, a
level surface of at least 50 cm x 60 cm with a load-bearing capacity of 80 kg
is recommended.
Installation conditions
►
Always set up the
MV-R 100
in a dry room that is as dust-free as
possible and ensure that no liquids can come into contact with the
device.
►
There must be no easily ignitable or flammable gases or liquids stored
in the installation rooms.
►
Do not put any flammable or ignitable objects in the vicinity of the
MV-R 100
.
►
Keep a minimum gap of 100 mm around the
MV-R 100
to ensure
adequate cooling.
Caution
Risk of ignition!
Ignition of the floor covering due to product tray breaking.
h
Ensure that the floor covering does not have a flammable surface.
Caution
Tipping loads!
Inadequate load-bearing capacity of the installation surface.
Ensure that the surface has adequate load-bearing capacity when setting
up the
MV-R 100
.
Caution
Risk of injury due to oven weight!
Physical overload/back problems due to heavy intrinsic weight.
h
The
MV-R 100
must be carried/moved by at least four people.
h
Mechanical strap eyelets are provided on the bottom of the device so
that it can be carried more easily.
Caution
Risk of overheating!
Overheating due to blocked air vents.
h
Ensure that the ventilation slots remain free on all sides.
Summary of Contents for 7203000100
Page 1: ...MADE IN GERMANY BETRIEBSANLEITUNG MV R OPERATING MANUAL...
Page 2: ......
Page 45: ......
Page 46: ......
Page 89: ......