background image

MM4300A

ISOLATED DC TO DC

TRANSMITTER

ADJUSTMENTS

The isolated transmitter is precisely
calibrated at the factory and does not
normally require user calibration. If
there is a need to recalibrate, proceed
as follows:

Connect a calibrated source to the
module input 

[terminals 5 (+) and               6

(-)]

.

Monitor the output of the module

[terminals 7 (+) and 8 (-)] 

with an

accurate digital meter.

Set the input signal to its low end, or
zero, value and adjust the ZERO control
to the proper low end output.

Increase the input signal to its full
scale value and adjust the SPAN
control for the high end  output signal.

Repeat the procedure once or twice,
the controls may interact slightly.

FUNCTION

The MM4300A provides a DC output,
fully isolated from input, line power
and ground, proportional to a DC
input. It is useful in eliminating ground
loops and common mode signals.

The module includes filtering and
conditioning to reduce susceptibility
to transients and noisy operations.
They utilize pulse width modulation
to develop a pulse train with a duty
cycle proportional to the input signal
amplitude. This pulse train is coupled
through a pulse transformer where
the duty cycle data is converted to a
proportional DC level in the output
circuit.

TYPICAL APPLICATIONS

Eliminates ground loops and large
common mode signals. Use for
v o l t a g e / c u r r e n t   s c a l i n g   a n d
conversion with isolation for buffering
and noise reduction.

CALIBRATION CONTROLS

The MM4300A Isolator contains two
calibration controls, zero and span
(gain). These controls are accessed
through the holes at the top of the
Mighty Module case, and are screw
driver adjusted.

ISOLATION, OUTPUT/INPUT

BREAKDOWN

>500 megohms,
>1000 VAC rms

BREAKDOWN,

POWER / CIRCUITRY

>1500 VAC rms

OPERATING TEMPERATURE

14°F to 140°F/
-10°C to 60°C

TEMPERATURE STABILITY

±0.02% of span
+30 µV/°C max

POWER

115 VAC ±10%,
50 or 60 Hz (2.5 W max)

24 VAC ±10%,
50 or 60 Hz (2.5 W max

230 VAC ±10%,
50 or 60 Hz (2.5 W max)

24 VAC ±10%,
50 or 60 Hz (2.5 W max)

(DC Power Option)

12 VDC
(limits 10 VDC to 15 VDC)
(2.5 W max)

24 VDC
(limits 21 VDC to 32 VDC)
(2.5 W max)

*Compliance:

The Sum of all voltage drops in the
output loop cannot exceed 24 V at
rated current (1200 ohms @ 20 mA).

SPECIFICATIONS

INPUT RANGES

4/20mA or 0/10VDC

INPUT IMPEDANCE

Voltage

200 kilohms

Current

50 ohms

OUTPUT RANGE

Voltage

0/+10 VDC,
2 kilohms minimum load

Current

4/20 mA,
24 V compliance*

RESPONSE TIME

£100 ms

OUTPUT RIPPLE

(peak-to-peak)

<0.1% of span

ACCURACY

±0.1% of span

LINEARITY

±0.05% of span

COMMON MODE REJECTION

120 dB, DC to 60 Hz

1

2

Reviews: