Height
117.5 cm
Width
90 cm
Depth
Depth with door open
70 cm
126.5 cm
Weight (net)
180 kg
Voltage, rated load, fuse
see data plate
Power cord
approx. 1.8 m
Water temperature: Cold water
Hot water and DI water connection
max. 20 °C
max. 70 °C
static water pressure
max. 1000 kPa overpressure
Min. flow pressure:
Hot and cold water
DI water connection
150 kPa overpressure
60 kPa overpressure
recommended flow pressure:
Hot and cold water
DI water connection
0
250 kPa overpressure
0
150 kPa overpressure
pressureless DI water connection
optional, 8.5 - 50 kPa
Drain height
min. 0.6 m, max. 1 m
Steam condenser
approx. 2 - 4 l/min
Room temperature
5 °C to 40 °C
Relative humidity maximum
linear decreasing to
80 % for temperatures up to 31 °C
50 % for temperatures up to 40 °C
Altitude above sea level
up to 1500 m
#
Contamination level
P2 (as per IEC/EN 61010-1)
Ingress protection (as per IEC 60529)
IP20 (ingress of dust)
Noise level in dB (A),
peak LpA for cleaning and drying cycles
< 70
Certification marks
CSA, Health Canada reg.Medical
device Class II, VDE, radio
interference protection
!
mark
MPG Guidelines - 93/42/EEC, Class IIb
Address of manufacturer
Miele & Cie. KG, Carl-Miele-Str. 29,
33332 Gütersloh, Germany
#
If the machine is installed above 1500 m, the washwater will boil at a lower tem-
perature. This may require you to lower the disinfection temperature and lengthen
the holding time (A
0
value).
Technical Data