Spring-loaded adapter for the
water supply
The spring-loaded adapter for the water
supply must be properly engaged
when a basket or the injector unit is
inserted into the washer-disinfector. In
other words, the adapter must be
4 - 5 mm higher than the water outlet in
the washer-disinfector.
If this is not the case, the adapter
needs to be adjusted as follows:
^
Loosen the locking ring
a
.
^
Move the adapter up to the proper
height
b
.
^
Tighten down the locking ring
c
.
How to adjust the upper basket
The height of the upper basket can be
adjusted on three levels, by 2 cm each.
Depending on the height selected for
the upper basket and which inserts are
used, labware/instruments of various
heights can be arranged in the baskets.
On the right-hand side, the height of
the upper basket is restricted by the
connector for the salt container.
To adjust the upper basket:
^
Pull the upper basket out as far as it
will go, then lift it up off the runners
and remove it.
^
Use a Torx T20 screw driver to
unscrew the roller bearings on either
side of the basket, then reposition as
desired.
Application
35