en - Areas of application
36
Butyrometers
The A 301/4 or A 301/5 modules are recommended for reprocessing
butyrometers.
The modules must be equipped with the SD-B nozzles for repro-
cessing butyrometers.
Due to the height of the nozzles, the module may only be used in the
A 150 lower basket. It is not possible to use an upper basket at the
same time.
A 302/2 and
A 302/3
The A 302/2 and A 302/3 modules each have 32 positions for repro-
cessing load items. The modules are designed for load items with a
volume from 20 ml to 100 ml.
Capacity
Volume
[ml]
Laboratory
flasks
Round
flasks
Erlenmeyer
flasks
Measuring
flasks
20–50
max. 32
max. 32
max. 32
max. 32
100
max. 32
1)
max. 16
2)
max. 16
2)
max. 16
2), 3)
1) the A 862 holding rack cannot hold any 100 ml laboratory flasks
2) plus 16 x load items with a smaller volume
3) A 843 injector nozzle required