en - FM flow meter and CM conductivity meter
38
CM Conductivity meter (full demineralisation)
A conductivity module
measures the conductivity level of the dem‐
ineralised water in a range from 0 - 199.9
µ
S/cm (microsiemens per
centimetre) and displays this in the conductivity meter
. This ena‐
bles quality control of the demineralised water.
If the defined limit value is reached, or if a fault occurs, an optical LED
lights up
and an acoustic signal is given.
If the demineralisation cartridge is depleted it must be replaced with a
new one and the spent resin sent away for reprocessing.
–
VE P 2000 and VE P 2800 cartridges:
Send cartridges filled with spent resin away for reprocessing.
–
LP 2800 cartridge:
Replace the resin. The E 316 Resin replacement set is supplied with
its own instruction manual.