Cleaning or replacing filters
Cleaning the
A 800 tubular filter
The tubular filter must be cleaned
at least once a week
.
Follow the cleaning instructions in the operating instructions for the
tubular filter.
Replacing FP 20
filter plates
Borosilicate filter plates must be replaced
approx. every 2 weeks
or
after 20 cycles
have been run.
To do this, unscrew the adapter.
Unscrew the top part from the bottom part.
Remove the silicone adapter
and the filter plate
from the bot-
tom part.
Rinse out the bottom part.
Insert a new FP 20 filter plate into the bottom part. The plate must
rest fully on the ring-shaped support of the bottom part.
Damaged or used filter plates must not be inserted under any
circumstances.
Insert the silicone adapter
into the bottom part.
Screw the top part and bottom part together.
Screw the adapter back onto the injector manifold.
Do not insert a filter plate into the holder if the injector manifold has
been fitted with an A 800 tubular filter.