Press
Clear (F4)
to erase all or selected test results.
Press the
ESCAPE
key to return to the
EXTENDED FUNCTIONS
screen.
See Chapter 3: Autotest for more information.
Autotest Setup
To display the
AUTOTEST SETUP
screen, press the
Extended Functions
key, then select
Setup
Autotest
. This screen lets you select the default Autotest, edit Autotest settings and delete user-
defined Autotests.
Press the
Ñ
or
Ò
arrow key to scroll through the list of Autotests (or to page up and
Ð
to
page down). When the Autotest you want is highlighted, press
Default (F1)
to select it as the
default Autotest.
Note: An Autotest name proceeded by an asterisk (*) indicates that the original Autotest has been
modified and the Autotest name has not been changed.
Press
Edit (F2)
to change the Autotest name, to select pairs to be used, to select the tests to run,
and to set test limits.
Press
Delete (F3)
to delete the highlighted user-defined Autotest.
Press the
ESCAPE
key to return to the
SETUP
screen.
See Chapter 3: Autotest for more information.
Editing Autotest Information
To display the
EDIT AUTOTEST
screen, press the
Extended Functions
key, then
select Setup
Autotest
, then press
Edit (F2).
This screen lets you customize Autotest settings.
Press the
Ñ
or
Ò
arrow key to scroll through the available options and press
Select (F1)
to
choose the highlighted option. Or, press the number on the keypad that corresponds to the option
you want.
•
Select
Name/Options
to edit the Autotest name and to change the Margin Options.
•
Select
Pairs
to choose the pairs to be tested during Autotest, and to indicate which
pairs have PASS/FAIL criteria applied to them during the Autotest.
•
Select
Tests to Run
to choose the tests to run during Autotest.
•
Select
Limits
to edit test limits.
•
Press the
ESCAPE
key to return to the
AUTOTEST SETUP
screen.
Calibrating the Injector
The Injector must be calibrated each calendar day that you use it to ensure accurate measurements.
To calibrate the Injector, connect the Scanner and the Injector with the patch cable supplied with
the Scanner. Press the
Extended Functions
key; then select
Calibrate Injector
.
Press
Cancel (F1)
or the
ESCAPE
key to stop the calibration.
See Chapter 2: Getting Started for more information on the Injector.